The blood of the common carotid artery (CCA) is directed to the ICA via a shunt that is commonly used in patients with contralateral carotid artery stenosis or circle of Willis disease. Monitoring with NIRS is an easy, noninvasive method for measuring continuous regional cerebral oxygen saturation (rSO2),[4] and the cost of the essential sensor needed to perform this procedure costs approximately 350 TL per patient ($210 USD).
With a cerebral oximetry device, beams from a light source are placed on two forehead electrodes, and these go through a banana-shaped path and return to the photo detectors on the electrodes. The ones close to the light source detect reflections from the skin, subcutaneous fat tissue, and bone tissue while those far from the light source detect the reflections from a deep watershed area at the junction of the anterior cerebral artery and the middle cerebral artery in the brain tissue. Blood flow rates in this region of the brain are 30% arterial and 70% venous; therefore, the values revealed by the cerebral oximetry devices show a mixture of arterial and venous blood at a ratio of 3:7.[5] Additionally, NIRS monitoring allows for the direct measurement of SO2, which is affected by the presentation and consumption in the cerebral cortex. Having this information could possibly give an early warning regarding perfusion defects. Furthermore, rSO2 monitoring has a low false-negative rate (2.6%) but a high false-positive rate (66.7%).[6]
TECHNIQUES
In this article, we aimed to present our clinical
experience related to the use of cerebral oximetry
monitoring method in the decision of shunt
usage in carotid surgery. In six of the 25 NIRSmonitored
patients (4 males, 2 females; mean age
70.3 years; range 67 to 75 years), a significant
decrease in cerebral SO2 was observed when the
cross-clamp was put into place. In our clinic,
following electrocardiography (ECG) and peripheral
oxygen saturation (SpO2) monitoring vascular access
was achieved with two 18 gauge cannula and
arterial monitoring was performed on patients who
would later undergo a CAE. Those with electrodes
placed on the right and left part of the forehead
for NIRS monitoring (INVOS® 5100C Cerebral/
Somatic Oximeter, Somanetics Corporation, Troy,
MI, USA) had surgery under general anesthesia after
recording their base values, and they were intubated with 1.2 mg/kg-1 rocuronium following anesthesia
induction with 2 mg/kg-1 fentanyl and 3-5 m g/kg-1
sodium thiopental.
In six of the 25 NIRS-monitored patients, a significant decrease in cerebral SO2 was observed when the cross-clamp was put into place, but the surgical team was warned that this would occur. Furthermore, increased cerebral oximetry values were observed after the use of a shunt in these six cases, and the data from these patients was recorded as follows: the mean right carotid stenosis was 75% (range 60-90%), the mean left carotid stenosis was 70% (range 60-80%), the mean operation time was 104.16 minutes (range 90-120), and the mean cross-clamp time was 27.16 minutes (range 25-32). Their preoperative and intraoperative NIRS values are presented in Table 1. We also observed that the cerebral oximetry values of the patients with unilateral or bilateral carotid stenosis were close to the normal values. However, the baseline values were a bit lower.
Table 1: Pre- and intraoperative near infrared spectroscopy values of the patients
The cerebral arterial system is correlated with the circle of Willis which receives blood from the four main arteries. When the internal and external carotid arteries are both cross-clamped on the operated side, we believe that the bilateral decrease of saturation is connected to the decrease in pressure throughout the whole system, especially in the early episodes.
In patients who do not require a shunt, the mean rSO2 difference was 10% of the baseline value, but this was not considered taken into account because 20% decreases from the baseline value are important in cerebral oximetry monitorization.
Carotid artery stump pressure is cheap and can be used to measure the ICA pressure after the clamping of the external artery and CCA. This measurement roughly assesses the adequacy of cerebral collateral circulation, but the lack of a standard procedure along with the accompanying multiple variables are the disadvantages of CASP. In addition, it does not allow for continuous monitoring.[4]
In a prospective study composed of 40 patients, Fassiadis et al.[7] investigated the correlation between NIRS monitoring and TCD in patients who underwent CEA under local or general anesthesia. They performed a correlation analysis regarding the percentage of decline in the mean flow velocity (FVm) between TCD and rSO2 via NIRS monitoring and reported that the rSO2 was well correlated with TCD. However, they determined that their inability to obtain reliable FVm readings on TCD in 35% of the patients was a serious disadvantage and that NIRS monitoring was a high-level device that provides a reliable and satisfactory evaluation of cerebral perfusion and oxygenation. In another prospective study by Pugliese et al.[1] that included 40 patients, the authors sought to determine the timing of the detection of an adverse event as well as the presence of false positivity and negativity. Therefore, they evaluated the neurological events and changes in parameters after carotid clamping utilizing TCD and rSO2 measurements and concluded that rSO2 was more reliable than TCD. In a different study conducted on 24 patients, Takeda et al.[8] detected cerebral ischemia with NIRS monitoring during a balloon occlusion test in conjunction with CEA and reported that a 10% decline from the baseline value was dangerous, but a 5% decline was safe.
The cerebral oximetry device is used in our hospital primarily in cases of acute dissection because of the limited number of electrodes. The useful information and postoperative data that we obtained from this device led us to also use it in carotid surgery. While shunt usage in this type of surgery is usually a decision of the surgeon after the stump pressure is measured in suspected cases, because a correlation was observed between the cerebral oximetry values and the stump pressure in our first 10 patients, the data we obtained from the NIRS monitoring provided the surgical team with the necessary confidence to consider the routine measurement of the stump pressure to be unnecessary. Thereafter, our surgical team followed our warnings after clamping the artery, and NIRS monitoring has now gained the trust of the surgical team. In addition, similar to our findings, Kragsterman et al.[9] reported that there was also a significant correlation between stump pressure and rSO2 values.
Declaration of conflicting interests
The authors declared no conflicts of interest with
respect to the authorship and/or publication of this
article.
Funding
The authors received no financial support for the
research and/or authorship of this article.
1) Pugliese F, Ruberto F, Tosi A, Martelli S, Bruno K,
Summonti D, et al. Regional cerebral saturation versus
transcranial Doppler during carotid endarterectomy under
regional anaesthesia. Eur J Anaesthesiol 2009;26:643-7.
2) Beneficial effect of carotid endarterectomy in symptomatic
patients with high-grade carotid stenosis. North American
Symptomatic Carotid Endarterectomy Trial Collaborators. N
Engl J Med 1991;325:445-53.
3) MRC European Carotid Surgery Trial: interim results for
symptomatic patients with severe (70-99%) or with mild
(0-29%) carotid stenosis. European Carotid Surgery Trialists\'
Collaborative Group. Lancet 1991;337:1235-43.
4) Cuadra SA, Zwerling JS, Feuerman M, Gasparis AP,
Hines GL. Cerebral oximetry monitoring during carotid
endarterectomy: effect of carotid clamping and shunting.
Vasc Endovascular Surg 2003;37:407-13.
5) Toroman F. The role of monitorization of regional cerebral
oxygen saturation in adult cardiac surgery. Journal of Cardio-
Vascular-Thoracic Anaesthesia and Intensive Care Society 2010;16:82-95.
6) Samra SK, Dy EA, Welch K, Dorje P, Zelenock GB,
Stanley JC. Evaluation of a cerebral oximeter as a monitor
of cerebral ischemia during carotid endarterectomy.
Anesthesiology 2000;93:964-70.
7) Fassiadis N, Zayed H, Rashid H, Green DW. Invos
Cerebral Oximeter compared with the transcranial
Doppler for monitoring adequacy of cerebral perfusion
in patients undergoing carotid endarterectomy. Int Angiol 2006;25:401-6.