Methods: Between February 2000 and February 2012, 1800 patients underwent pediatric cardiac operation in our department. Among these patients 10 (5 boys, 5 girls; median age 3 months; range 2.5 month to 4.5 years) needed mechanical assist devices for postcardiotomy cardiac failure and cardiopulmonary failure in the intensive care unit. In four patients, roller pump support with left atrial appendage and ascending aorta cannulation was used for temporary left ventricle support. In six patients, extracorporeal membrane oxygenation (ECMO) was used for temporary cardiac and pulmonary support.
Results: Three patients in the roller pump support group (75%) and three patients in the ECMO support group (50%) were successfully weaned from the mechanical assist. The longest mechanical support with the roller pump for left ventricle dysfunction was 160 hours in a seven-month-old infant after arterial switch operation. Mechanical support was required in two patients in the roller support group including one (4 years old) after arterial switch operation and one (2.5 months old) due to the left ventricular dysfunction after surgery for anomalous left coronary artery from pulmonary artery. Both patients were successfully weaned from the roller pump support after 48 hours. In two patients ECMO support was necessary for cardiac arrest due to pulmonary hypertensive crisis in the intensive care unit following the ventricular septal defect closure and ECMO was successfully weaned off after 12 and 48 hours.
Conclusion: Pediatric mechanical circulatory support is a life-saving procedure for patients who develop cardiopulmonary failure after cardiac surgery and are unresponsive to the conventional resuscitative efforts. Conventional roller pump support is an alternative method for temporary left ventricle dysfunction in selected cases.
When cardiopulmonary failure occurred in the operating room, ECMO support was initiated with right atrial and aortic cannulae, whereas left VAD support was started with left atrial appendage and aortic cannulation (Medtronic DLP arterial cannulae 8-12F and Medtronic DLP venous metal angled cannulae 14-24F. Medtronic Inc., Minneapolis, MN, USA) for left ventricle failure in the operating room. A left atrial catheter or a pulmonary artery catheter along with atrial and ventricular pacing wires were then inserted. Sternum was not closed, and the mediastinum was covered with a plastic membrane. For cardiopulmonary failure in the ICU, support with a standard CPB system was started, and ECMO system was replaced, or ECMO support was started directly in the ICU with Medtronic ECMO oxygenator system (Medtronic Inc., Minneapolis, MN, USA). In addition, a Terumo Sarns roller pump (Terumo Medical Corporation, Somerset, NJ, USA) was used for left ventricle support. Pump flow was adjusted to full or 75% flow maintaining left atrial pressures of 10-15 mmHg in the first two days. Gradual weaning was planned during this period depending on left atrial pressure, blood gasses, pulmonary indices, and ventricular function [shortening fraction (SF) >28%] with serial echocardiographic examinations.
Anticoagulation was maintained with heparin infusion and activated clotting time (ACT) monitoring. The target ACT for the left VAD was 130-180 seconds while the target ACT for ECMO was 180-220 seconds. Blood products were transfused to maintain hematocrit levels of between 35 and 45% and a platelet count of >50.000/mm3. Furthermore, albumin and fresh frozen plazma (FFP) were transfused, and low-dose inotropic and vasopressor support, furosemide infusion, peritoneal dialysis, and ultrafiltration were also employed to maintain fluid balance. The circuit pressure was constantly checked, and the patient was continuously monitored by the perfusionists for mechanical device failure, air bubbles, and hemolysis. The patient’s lactate dehydrogenase level and urine analysis were used for monitoring the hemolysis. In addition, the raceway tubing was replaced every 72 hours and repositioned every 24 hours to prevent the risk of rupture. ECMO weaning and decannulation was routinely performed in the operating room.
Recoverable cardiopulmonary dysfunction and bridging for transplantation are the two main indications for the need for mechanical assist devices in the pediatric age group. The decision is made according to the patient’s clinical condition, technical and logistical sources, and clinical experiences. Conventional CPB systems, ECMO systems, Berlin Heart EXCOR® p ediatric V AD ( Berlin H eart, GmbH, The Woodlands, TX, USA), Medos VAD (Medos Medizintechnik AG, Stolberg, Germany) and centrifugal pumps are among the currently available support devices. In 1973, Soeter et al.[2] reported the first use of extracorporeal support for a four-year-old child after tetralogy of Fallot (TOF) repair when respiratory failure developed. After 48 hours of support, the child was successfully weaned. Furthermore, Barlett et al.[3] reported the results of the clinical use of ECMO in a case of cardiopulmonary failure in 1976 and 1977. The Extracorporeal Life Support Organization (ELSO) was also established in 1989 and began to collect ECMO data from related centers.
In addition to the conventional CPB and ECMO systems, there are two types of mechanical assist devices: intracorporeal and paracorporeal. Berlin Heart EXCOR® pediatric VAD, Medos VAD, Thoratec VAD, and the AbioMed BVS 5000/AB 5000 (AbioMed, Inc., Danvers, MA, USA) are paracorporeal devices in which the pump, connections, and console are outside of the body. In intracorporeal devices like the DeBakey VAD® Child (MicroMed Technology, Inc., Houston, TX, USA), Jarvik 2000 VAD (Jarvik Heart, Inc., New York, NY, USA), Berlin Heart INCOR® (Berlin Heart, GmbH, The Woodlands, TX, USA), and Thoratec Heartmate II (Thoratec Corporation, Pleasanton, CA, USA), the pump and main connections are inside the body. However, many of these devices are not suitable for children today, and the use of intraaortic balloon pump support is limited in pediatric cardiac surgery because of technical considerations and the elasticity of the aorta.
Two support modalities are recommended for pediatric cardiac patients: ECMO, which supports both pulmonary and cardiac functions, and left and right VAD systems, which support only cardiac (left and/or right ventricle) functions.[4] The Berlin Heart EXCOR® is currently the most widely used pediatric left/right VAD system.[5] Roller pump support for temporary left ventricular failure was the choice in our patients because of the easy set-up and lack of available pediatric centrifugal pumps or other left VAD systems. Fortunately, new support systems have recently become available in our country, and we aim to use these novel devices for our future patients.
Support systems are most frequently used in pediatric cardiac surgery when there is difficulty in weaning from CPB, low cardiac output, severe left, right ventricular or biventricular failure, respiratory failure, or postoperative cardiopulmonary arrest. However, severe neurological injuries, irreversible sepsis, multiorgan failure, severe coagulation disorders, multiple congenital anomalies, and extreme prematurity are contraindications for mechanical support. The most common complications are bleeding, thromboemboli, hemolysis, infection, and multi-organ failure.[6] Bleeding was decreased in our last two patients via roller pump left VAD support, with for whom we maintained ACT levels of between 120 and 150 seconds, and the use of tubing sets with Terumo Xcoating™ (Terumo Medical Corporation, Somerset, NJ, USA). We did not observe significant hemolysis with negative clinical outcomes in our roller pump-supported patients.
Before starting the weaning process, it is important to evaluate the echocardiographic findings. In our patients, weaning process was started when there was an improved SF contractility of >28%, a mean arterial pressure (MAP) of >50 mmHg, and a left atrial (LA) pressure of <15 mmHg. The support level was decreased by increments of 25% every four-six hours, and once it reached a level of 25%, the patients were transferred to the operating room for safe cannullae removal. Support management for ECMO is more difficult than for VADs and has a higher risk of bleeding, thrombosis, embolization, major organ dysfunction (pulmonary, neurological, renal, and gastrointestinal), and infection. We have observed that surgeons, anesthetists, intensivists, and perfusionists along with ICU and operating room nurses need advanced training to establish and maintain a successful ECMO program. In our cases that involved cardiopulmonary failure, we used ECMO (Figure 1), and for left ventricle failure, we chose the roller pump (Figure 2). Short-term centrifugal pump support and long-term support with the Berlin Heart EXCOR® might be better alternatives for children, and we plan to use these devices in the future.
Figure 1: Extracorporeal membrane oxygenation support in the intensive care unit.
Figure 2: Left ventricular assist device support with roller pump in the intensive care unit.
Pediatric mechanical cardiopulmonary support is a life saving procedure for patients who develop cardiopulmonary failure after cardiac surgery and do not respond to conventional resuscitative efforts.[7-10] It has always been difficult to manage bleeding problems in these patients, but new ECMO systems need a lower degree of anticoagulation. Centers who deal with these patients should have a blood bank and blood product support and be able to provide the specific laboratory tests that are needed. Certain groups of patients, such as those with anomalous origin of the left coronary artery from the pulmonary artery (ALCAPA) pathology and those for whom an arterial switch operation with left ventricle involution is planned, might be candidates for elective ECMO and left VAD support. Customized support systems for the right and left ventricles, biocompatibility, anticoagulation, control systems, size, price, reliability, and durability are all issues that still need improvement in pediatric cardiopulmonary support systems.[11-14]
One limitation of our study was that, it was made up of a small group of patients; therefore, it was difficult to draw definitive conclusions. In addition, it was not possible to compare our results because there were no available alternative support systems. Nevertheless, we believe that our experiences with these patients could prove to be valuable for centers in countries with limited resources.
Declaration of conflicting interests
The authors declared no conflicts of interest with
respect to the authorship and/or publication of this
article.
Funding
The authors received no financial support for the
research and/or authorship of this article.
1) Erek E, Yalcınbaş YK, Sarioglu T. Acute aortic insufficiency
and its surgical treatment early after arterial switch operation:
a report of two cases. Turk Gogus Kalp Dama 2009;17:117-9.
2) Soeter JR, Mamiya RT, Sprague AY, McNamara JJ.
Prolonged extracorporeal oxygenation for cardiorespiratory
failure after tetralogy correction. J Thorac Cardiovasc Surg
1973;66:214-8.
3) Bartlett RH, Gazzaniga AB, Fong SW, Jefferies MR, Roohk
HV, Haiduc N. Extracorporeal membrane oxygenator
support for cardiopulmonary failure. Experience in 28 cases.
J Thorac Cardiovasc Surg 1977;73:375-86.
4) Cooper DS, Jacobs JP, Moore L, Stock A, Gaynor JW,
Chancy T, et al. Cardiac extracorporeal life support: state of
the art in 2007. Cardiol Young 2007;17 Suppl 2:104-15.
5) Potapov EV, Stiller B, Hetzer R. Ventricular assist devices
in children: current achievements and future perspectives.
Pediatr Transplant 2007;11:241-55.
6) Hehir DA, Niebler RA, Brabant CC, Tweddell JS, Ghanayem
NS. Intensive care of the pediatric ventricular assist device
patient. World J Pediatr Congenit Heart Surg 2012;3:58-66.
7) Tajik M, Cardarelli MG. Extracorporeal membrane
oxygenation after cardiac arrest in children: what do we
know? Eur J Cardiothorac Surg 2008;33:409-17.
8) Alsoufi B, Al-Radi OO, Gruenwald C, Lean L, Williams
WG, McCrindle BW, et al. Extra-corporeal life support
following cardiac surgery in children: analysis of risk factors
and survival in a single institution. Eur J Cardiothorac Surg
2009;35:1004-11.
9) Naim MY, Topjian AA, Nadkarni VM. CPR and E-CPR:
What is New? World J Pediatr Congenit Heart Surg 2012;3:48-53.
10) Connelly JT, Weaver B, Seelhorst A, Beaty CD, McDonough
M, Nicolson SC, et al. Challenges at the Bedside With
ECMO and VAD. World J Pediatr Congenit Heart Surg
2012;3:67-71.
11) Shum-Tim D, Duncan BW, Hraska V, Friehs I, Shin’oka T,
Jonas RA. Evaluation of a pulsatile pediatric ventricular
assist device in an acute right heart failure model. Ann
Thorac Surg 1997;64:1374-80.
12) Pekkan K, Frakes D, De Zelicourt D, Lucas CW, Parks WJ,
Yoganathan AP. Coupling pediatric ventricle assist devices to
the Fontan circulation: simulations with a lumped-parameter
model. ASAIO J 2005;51:618-28.