The patient was admitted to the catheter laboratory with the aim of device closure of the tunnel using transesophageal echocardiography (TEE) guidance. Bilateral femoral artery access was obtained, and 100 units/kg of heparin were administered. An ascending aortogram performed in the right anterior oblique and lateral views using a 6F pigtail catheter revealed a sigmoid-shaped tunnel measuring 5.8 mm at the aortic orifice, 7.8 mm in the middle section and 11 mm at the left ventricular opening (Figure 2). After crossing the defect with 0.018” Pointer® guidewire through a 6F right Judkins catheter (Medtronic Inc., Minneapolis, MN, USA), balloon interrogation with an 8x30 mm Tyshak® balloon was performed to delineate the tunnel’s size and course. The defect showed a loose indentation on the balloon at its aortic opening. Then, an appropriate guiding catheter was advanced into the LV over the wire. A 12 mm AVP II (St. Jude Medical, Inc., St. Paul, Minnesota, USA) was utilized initially and placed in the middle section of the defect. However, it could not be released due to the significant residual regurgitant flow seen on the control aortogram. A 14x12 mm Amplatzer duct occluder (ADO) then was implanted into the tunnel. The device occluded the defect completely but caused significant aortic insufficiency (AI). Ultimately, it was only possible to occlude the defect with a 16 mm AVP II without significant AI on TEE.
Figure 2: Right anterior oblique aortagram revealed a sigmoidshaped tunnel (arrows).
After releasing the device, the control angiogram showed complete occlusion of the tunnel (Figure 3, 4). The coronary arteries were patent, and there was no demonstrable aortic regurgitation. The following day, echocardiography also showed complete occlusion of the defect with trivial AI (Figure 5, 6). The patient was discharged 24 hours after the procedure and started aspirin at a dose of 5 mg/kg/day. Eight months after the device closure, echocardiogram showed that the device was positioned well with no residual shunt. There was trivial aortic regurgitation with normal LV end diastolic dimension and systolic function (LVEDd: 52 mm, LVEDd Z-score: +1,27, EF: 58%, FS: 29%).
Figure 5: Transthoracic parasternal long axis view revealed complete occlusion of tunnel.
Figure 6: Transthoracic parasternal short axis view revealed complete occlusion of tunnel.
Major concerns about the transcatheter device closure include proximity of the defect to the coronary arteries, compression and deformation of the aortic valve, and incomplete closure that can cause significant hemolysis.[3,4,8] However, we believe that device closure of ALVT using an AVP II has several advantages over surgery. The procedure eliminates the need for sternotomy and cardiopulmonary bypass and shortens hospital stay. It also precludes surgeryrelated complications, such as impingement of the right coronary cusp, distortion of the aortic valve and left bundle branch block that occurs due to sutures placed in the aortic and ventricular openings. Regardless of surgical technique, researchers have reported an incidence of AI in patients with ALVT after surgical closure up to 60%.[9] On the contrary, the AVP II may act as a supporting tissue to the right coronary cusp by filling the gap created by the tunnel, thus avoiding future aortic valve dysfunction.
The AVP II is a cylindrical self-expanding device made of nitinol wire mesh and indicated for peripheral arterial and venous embolization. It has a central lobe surrounded by two shorter occlusion discs. These three lobes are all of the same diameter and create six planes of occlusion.[10] The manufacturer recommends choosing an AVP II size 20 to 50% larger than the target vessel diameter for complete occlusion. However, in their study on 178 pediatric patients with patent ductus arteriosus, Garay et al.[10] needed an AVP II size 2.6 times the ductus’ narrowest diameter for satisfactory occlusion. In our patient, a 12 mm AVP II could not completely occlude the tunnel, and the size of final implanted device was 2.7 times the narrowest tunnel diameter. Therefore, we believe that larger-thanrecommended sizes of AVP II may be required in these distensible defects.
Use of AVP II in ALVT has several advantages over ADO I and other devices with a retention disc larger than the main body. The larger retention disc in these devices may cause hemolysis due to shear stress on the erythrocytes when remaining in the left ventricle.[4] As seen in our patient, the retention discs may distort the aortic valve and result in significant aortic insufficiency when left inside the tunnel. In addition, the proximal retention skirt on a double disc device may obstruct the coronary artery entries when left on the aortic side. In our patient, the AVP II, with its flexible and soft nature, perfectly conformed to the shape of the ALVT. Due to its design with three lobes of the same diameter, the AVP II also eliminated the risk of protrusion of the retention skirts into the left ventricular cavity or ascending aorta.
In conclusion, our report suggests that it is feasible to close an aortico-left ventricular tunnel via a transcutaneous procedure, depending on the size and morphology of the tunnel and the patient age. Amplatzer® Vascular Plug II may be preferable in transcatheter closure of aortico-left ventricular tunnel because of its soft structure and strong occlusive characteristics. However, the dimensions of aorticoleft ventricular tunnel measured on echocardiography and angiography may be misleading due to the complex morphology and distensibility of the defect; therefore, larger than predicted devices may be required.
Declaration of conflicting interests
The authors declared no conflicts of interest with respect to
the authorship and/or publication of this article.
Funding
The authors received no financial support for the research
and/or authorship of this article.
1) Mahle WT, Kreeger J, Silverman NH. Echocardiography
of the aortopulmonary window, aorto-ventricular tunnels,
and aneurysm of the sinuses of Valsalva. Cardiol Young
2010;20:100-6.
2) Vijayalakshmi IB, Chitra N, Prabhu Deva AN. Use of
an Amplatzer duct occluder for closing an aortico-left
ventricular tunnel in a case of noncompaction of the left
ventricle. Pediatr Cardiol 2004;25:77-9.
3) Thomas E, Maskari S, Farqani A. Percutaneous device
closure of aortico-left ventricular tunnel using Amplatzer
vascular plug III. Cardiol Young 2013;23:755-8.
4) McKay R. Aorto-ventricular tunnel. Orphanet J Rare Dis
2007;2:41.
5) Chessa M, Chaudhari M, De Giovanni JV. Aorto-left
ventricular tunnel: transcatheter closure using an amplatzer
duct occluder device. Am J Cardiol 2000;86:253-4.
6) Norman R, Kafka H. Aortico-left ventricular tunnel in adulthood:
twenty-two year follow up. Int J Cardiol 2009;134:20-2.
7) Kim RW, Spray TL. Surgical correction of aortico-left
ventricular tunnel. Semin Thorac Cardiovasc Surg Pediatr
Card Surg Annu 2006:177-9.
8) Martins JD, Sherwood MC, Mayer JE Jr, Keane JF. Aorticoleft
ventricular tunnel: 35-year experience. J Am Coll Cardiol
2004;44:446-50.