Fifty coronary artery patients who were operated electively In Ege University Medical Faculty Cardiovascular Surgery Department were taken to this study and randomized in two groups. BRAT (Baylor Rapid Auto Transfusion System) cell saver apparatus has been used in the first group. The results obtained from this study group have been compared with those obtained from the control group to find out if there were any statistical significance between them.
In this study, the total drainage and the number of blood transfusions in the first 24 hours and the amount of collected blood by cell saver were recorded. The radial artery blood samples were taken 3 times for each patient (preoperatively, postoperatively first and second days). Haematoc rit, activated clotting time, pro thrombin time, activated partial thromboplastin time, the levels of fibrinogen, total protein, albümin, C3, C4, and whit blood cells, red blood cells, platelets counts were measured in these samples.
The need for transfusion of blood and blood products during open heart surgery as well as C3 complement levels decreased significantly by using cell saver. Also the amount of drainage decre ased in this group but this is not statistically significaııt. No harmful side effects due to cell saving was observed.
Otolog kan transfüzyonu yapılma şekilleri arasında, elektif cerrahi girişimden önce günler, haftalar, hatta aylar süren kanın 'predonation'u[1,4,5,8,15], pre ya da perioperatif hemodilüsyon[1,15], çeşitli yöntemler ve cihazlarla kanın korunması ve hastaya geri transfüzyonu (ekstrakorporal dolaşım hatlarında kalan tüm volümün hastaya geri verilmesi, mediastinal drenlerden gelen kanın oto transfüzyonu, intraoperatif hemofiltrasyon ve hemokonsantrasyon, aprotinin vs. kullanımı)[1,4,7,9,10,17,18] sayılabilir.
Bu çalışmada BRAT (Baylor Rapid Autotransfu sion System) cell saver cihazı kullanılan hastalarla kullanılmayanlar karşılaştırılmış ve sonuçlar istatistiksel olarak değerlendirilmiştir.
Cell saver yönteminin en önemli dezavantaj larından birisi rutin kullanım için fiyatının oldukça pahalı olmasıdır[1,15]. Bu durum, cell saver'ın sadece seçilmiş vakalarda kullanılmasına yol açmaktadır. Bir diğer önemli dezavantajı ise, plazma proteinle rini, trombosit ve lökositleri de beraberinde yıkayıp kandan uzaklaşmalarına yol açtığı için teorik ola rak postoperatif kanama ve enfeksiyon riski oluş turduğu düşünülmesidir[10,11,15]. Ancak bu durum hiçbir çalışmada gösterilemediği gibi, bu çalışmada da enfeksiyon gözlenmemiş ve cell saver kullanılan grupta drenaj daha az bulunmuştur.
Çalışmada cell saver kullanımının istatistiksel olarak anlamlı biçimde kan kullanımını ve komple man aktivasyonunu azalttığı gösterilmiştir. Anlamlı olmasa da cell saver kullanılan grupta drenaj daha az bulunmuştur. Sonuç olarak, cell saver kullanımının homololg kan kullanımını önemli ölçüde azalttığı, kan proteinleri, trombosit sayısı ve diğer koagülasyon faktörlerinde önemli bir azalmaya neden olmadığı, dolayısıyla da ACT ve APTZ'yi uzatmadığı gözlenmiştir. Diğer yandan toksik metabolitlerin vücuttan atılmasında ve homolog kan kullanımının azaltılmasında son derece yararlı bir yöntemdir. Kanama riski olan ve kan temininde güçlük çekilen hastalarda güvenle kullanılabilir.
1) Ovrum E, Holen E, Abdelnoor M, Oystese R: Conventional Blood Conservation Techniques in 500
Consecutive Coronary Artery Bypass Operations.
AnnThorac Surg 52:500-5,1991.
2) Cosgrove DM, Loop FD, Lytle BW: Blood conservati
on in cardiac surgery. Cardiovasc Clin 12:165-75,
1981.
3) Schaff HV, Hauer JM, Bell WR, et al: Autotransfusion
of shed mediastinal blood after cardiac surgery. A
prospective study. J Thorac Cardiovasc Surg 75:632-
41,1978.
4) Love TR, Hendren WG, O'Keefe DD, Daggett WM;
Transfusion of predonated autologous blood in elective cardiac surgery. Ann Thorac Surg 43:508-12,
1987.
5) Britton LW, Eastlund DT, Dziuban SW, et al: Predo
nated autologous blood use in elective cardiac surge
ry. Ann Thorac Surg 47:529-32,1989.
6) Dietrich W, Barankay A, Dilthey G, et al: Reduction
of blood utilization during myocardial revasculariza
tion. J Thorac Cardiovasc Surg 97:213-9,1989.
7) Breyer RH, Engelman RM, Rousou JA, et al: A com
parison of Cell Saver versus ultrafilter during coro
nary artery bypass operations. J Thorac Cardiovasc
Surg 90:736-40,1985.
8) Russell GN, Peterson S, Harper SJ, Fox MA: Homolo
gous blood use and conservation techniques for car
diac in the United Kingdom. Bri Med J 297:1390-1,
1988.
9) Hartz RS, Smith JA, Green D: Autotransfusion after
cardiac operation. Asscssment of hemostatic factors.
J Thorac Cardiovasc Surg 96:178-82,1988.
10) Scott WJ, Kessler R, Wernly JA: Blood conservation
in cardiac surgery. Ann Thorac Surg 50:843-51,1990.
11) Collins JA: Problems associated with massive transfuisions ofstored blood. Surgery 75:274-95,1974.
12) Tennenberg SD, Clardy SW, Bailey WW, Solomkin
JS: Complement activation and lung permeability
during cardiopulmonary bypass. Ann Thorac Surg
50:597-601,1990.
13) Chcnoweth DE, Cooper SW, Hugli TE, et al: Complement activation during cardiopulmonary bypass:
evidence for genertion of C3a and C5a anaphylato
xins. N Engl JMed 304:497-503,1981.
14) Kirklin JK, Westaby S, Blackstone EH, et al: Complement and the damaging effects of cardiopulmonary
bypass. J Thorac Cardiovasc Surg 86:845-57, l 983.
15) Taylor KM: Blood conservation during cardiopulmo
nary bypass - autologous transfusion, cell saving and
haemofiltration. Cardiopulmoray bypas. Chapman
and Hall Ltd 17:289-307,1986.
16) Page PA: Ultrafiltration versus cell washing for
blood concentration. J Extracorp Tech 22:142-150,
1990.