ISSN : 1301-5680
e-ISSN : 2149-8156
Turkish Journal of Thoracic and Cardiovascular Surgery     
Use of Spectrum Medical M3 monitor in pediatric cardiac surgery: gaseous emboli monitoring in Capiox RX05 and Capiox FX05 oxygenators
Yusuf Kenan Yalçınbaş1, Serdar Günaydın2, Murat Boz1, Zekeriya Telli1, Tayyar Sarıoğlu3
Department of Cardiovascular Surgery, Acıbadem Bakırköy Hospital, İstanbul, Turkey
1Department of Cardiovascular Surgery, Acıbadem Bakırköy Hospital, İstanbul, Turkey
2Department of Cardiovascular Surgery, Medline Eskişehir Hospital, Eskişehir, Turkey
3Department of Cardiovascular Surgery, Medical Faculty of Acıbadem Hospital, İstanbul, Turkey
DOI : 10.5606/tgkdc.dergisi.2014.10444
Background: The clinical use of Spectrum Medical M3 monitor, and the results of emboli monitoring for Terumo Capiox RX05 oxygenator without arterial filter and Terumo Capiox FX05 oxygenator with built in arterial filter were evaluated.

Methods: Spectrum Medical M3 monitor was used for 50 patients between December 2010 and June 2011. Capiox RX05 oxygenator was used for 30 patients (group 1; 21 boys, 9 girls; mean age 26.3±12.4 months), and Capiox FX05 oxygenator was used for 20 patients (group 2; 10 boys, 10 girls; mean age 22.4±12 months). Heparin dose was 400 units/kg. Activated clotting time was maintained at >500 seconds, hematocrit was maintained at 25-30%, and cardiopulmonary bypass flow was maintained at 125-150 mL/kg/minute. Moderate hypothermia (28- 34 degrees), alfa stat blood gas management, and antegrade isothermic blood cardioplegia were used.

Results: No mortality was observed in group 1. In group 2, one patient died due to pulmonary hypertension. Two patients needed prolonged respiratory support for pneumonia in group 1 and group 2. In group 1, the mean number of arterial gaseous emboli was 32573 (23063-60051) (25-75% percentile), and mean volume of arterial gaseous emboli was 4.2 mL (2.1-7.5 mL) (25-75% percentile). In group 2, the mean number of arterial gaseous emboli was 25053 (9096.3-37352.3) (25-75% percentile), and mean volume of arterial gaseous emboli was 1.1 mL (0.5-3.2 mL) (25-75% percentile) (p value for t he number of a rterial emboli was 0.043, and p value for arterial emboli volume was 0.001).

Conclusion: Spectrum Medical M3 monitor is a useful device for monitoring gaseous emboli in pediatric cardiac surgery. Terumo Capiox FX05 oxygenator with built in arterial filter may be more effective for the prevention of gaseous emboli than Terumo Capiox RX05 oxygenator without arterial filter.

Keywords : Cardiopulmonary bypass; gaseous emboli; pediatric cardiac surgery
Viewed : 10257
Downloaded : 3325