Outflow graft twisting of Heartmate III left ventricular-assisted device: A case report
Oğuz Konukoğlu1, Denyan Mansuroğlu1, Özgür Yıldırım1, Shiraslan Bakshaliyev1, Kenan Sever1, Mehmet Balkanay1
1Department of Cardiovascular Surgery, Yeni Yüzyıl University, Private Gaziosmanpaşa Hospital,, Istanbul, Turkey
DOI : 10.5606/tgkdc.dergisi.2019.17871
Outflow graft twisting of Heartmate III left ventricular-assisted
device is a rare device complication and, due to the recent
reports from transplant centers, the firm and the United States
Food and Drug Administration have announced explanatory
measures. When the outflow graft twisting occurs, the diagnosis
and treatment should be done without any delay due to serious
decline in the device output and unstable hemodynamics. This
article describes the first case of an outflow graft twisting in
Turkey.
Keywords : Graft, Heartmate III, left ventricular-assisted device, outflow, twisting