Methods: This prospective, randomized study included a total of 40 patients (36 males, 4 females; mean age 60.0±8.5 years; range 43 to 75 years) who underwent elective coronary artery bypass grafting between January 2015 and December 2015. The patients were divided into two groups by block randomization using the sealed envelope technique. In group 1, non-integrated arterial filter membrane oxygenators and, in group 2, integrated arterial filter membrane oxygenators were used. Near-infrared spectroscopy was used to assess the cerebral oxygenation intraoperatively in all patients. Cranial diffusion-weighted magnetic resonance imaging was performed 2-4 days before and after the surgical procedure. Cognitive functions were evaluated using the Montreal Cognitive Assessment at the postoperative one month.
Results: Eleven patients in the non-integrated group and seven patients in the integrated group had new lesions in the diffusion-weighted magnetic resonance imaging. The mean pre- and postoperative total Montreal Cognitive Assessment scores were 27.9±3.3 vs 28.1±3.4 and 26.2±3.1 vs 26.8±3, respectively, in the non-integrated and integrated groups. There were no statistically significant differences between the two groups in terms of the number of new lesions, the near-infrared spectroscopy findings, and the Montreal Cognitive Assessment scores.
Conclusion: Membrane oxygenators with integrated arterial filters do not seem to offer a significant advantage over those without integrated arterial filters in terms of neurocognitive outcomes in patients undergoing coronary artery bypass grafting.
Several strategies have been developed to minimize the entry of micro-bubbles into ECC, and to decrease the surface area as well as the prime volume.[5,6] New generation oxygenators with integrated arterial filters which allow the incorporation of arterial filters into the oxygenators have been reported to decrease cerebral injury through entrapment of the particulate matter and micro-bubbles.[7,8] Integration of arterial-line filters with oxygenators as a novel CPB design was contrived for integrating the concepts of surface covering, blood filtration, and miniature.[9] However, there is a limited number of studies comparing conventional CPB equipment with oxygenators with integrated arterial filters.
In the present study, we aimed to prospectively compare integrated vs non-integrated arterial filters in terms of postoperative neurocognitive functions and neuroradiological changes in patients undergoing coronary artery bypass grafting (CABG).
A total of 58 patients who underwent elective CABG between January 2015 and December 2015 were included in the study. The patients were divided into two groups by block randomization using the sealed envelope technique. In group 1 (n=28), non-integrated arterial filter membrane oxygenators and, in group 2 (n=30), integrated arterial filter membrane oxygenators were used during CABG surgery. Exclusion criteria included an age less than 18 or older than 90 years, concomitant surgery in addition to CABG, history of cerebrovascular disease, poor educational status, presence of mental or psychiatric disorders, prolonged intensive care unit stay due to any reason, hemodynamic instability, and a carotid stenosis higher than 70% as documented by Doppler ultrasound. Cranial diffusionweighted magnetic resonance imaging (MRI) was performed two to four days before and after CABG. Preoperative cognitive test was administered within seven days prior to CABG, and postoperative cognitive tests were performed at one month postoperatively.
A total of 58 patients underwent cognitive testing prior to surgery. Five patients were excluded due to emergent chest pain, which required close clinical monitoring before surgery. Two patients did not have surgery. Six patients were unable to attend to postoperative MRI visits due to scheduling problems. Five patients refused the follow-up scan. Therefore, a total of 40 patients were prospectively analyzed.
Anesthesia and surgical procedure
All procedures were performed with a Sorin C5
roller pump (Sorin Group USA, Inc., Arvada, CO,
USA) using polyvinyl chloride tubing in both groups.
Dideco Evo non-integrated arterial filter membrane
oxygenators (Sorin Group Italia S.R.L., Milano, Italy)
were used in group 1, while Medtronic Affinity
(Medtronic, Inc. Minneapolis, USA) fusion integrated
arterial filter membrane oxygenators were used in
group 2.
All operations were performed through median sternotomy. For all patients, before sternotomy, intraoperative monitoring with transesophageal echocardiography was done to measure the aortic wall thickness and to detect atherosclerotic plaques and calcifications. The aortic wall thickness alterations were recorded according the classification defined by Katz et al.[10] The aortic wall thickness was defined as mild (intimal thickness <3 mm), moderate (intimal thickness 3-5 mm), and severe (intimal thickness >5 mm). Aortic calcifications were evaluated both by intraoperative transesophageal monitoring and manually. Patients who had calcifications and ulceration at the proximal anastomosis and cross-clamp site were excluded from the study.
After the administration of 300 U/kg heparin, CPB was established by aortic and right atrial two-stage single venous cannulation. An aortic root cannula was inserted into the ascending aorta for cardioplegia and venting. Moderate hypothermia (28 °C - 32 °C) was used. When necessary, additional doses of heparin were administered to maintain the activated clotting time over 400 sec. Cold blood cardioplegia (18 °C) was used to achieve cardiac arrest and intermittent antegrade myocardial protection. Partial arterial pressure of carbon dioxide (pCO2) was maintained at 35-40 mmHg, and the arterial tension of oxygen was maintained at 200-250 mmHg during CPB. Non-pulsatile blood flow during CPB was maintained at 2.4 L/min/m2 body surface area at normothermia and was adjusted as needed in case of hypothermia. The mean arterial pressure was kept above 55 mmHg throughout the operation using vasoactive agents. Also, packed red cells were administered, when necessary to sustain the hematocrit value above 21% during CPB. Alpha-stat protocol was used to manage the acid-base status.
Cognitive assessments
To assess the alterations in cognitive function, a
cognitive test was applied to all participants pre- and
postoperatively by an experienced neurologist. The
results were evaluated by two blinded observers.
Neurological assessments included the evaluation of
visual and speech disorders, paralysis in the upper
or lower extremities, or mental state disorders. The
Montreal Cognitive Assessment (MoCA) test was
used to identify cognitive alterations.[11] This tool
was originally introduced as a fast screening test for
mild cognitive disorder. It assesses the alterations in
attention and concentration, coordinating functions,
memory, language, visual structuring skills, abstract
thought, calculation and orientation. The maximum
attainable score is 30. Under these circumstances,
a score of ≥26 points indicates normal cognitive
functions; 19-25 indicates mild cognitive disorder,
and <19 indicates severe cognitive decline.[11,12]
Near-infrared spectroscopy (NIRS)
Cerebral oxygenation of patients was monitored
using the near-infrared spectroscopy during surgery.
Baseline values were calculated before the induction of
anesthesia, and the subsequent regional oxygen saturation
(rSO2) values were recorded after cross-clamping.
Decline in saturation was recorded as percentage,
and an intervention was performed when a decrease
higher than 20% occurred. The NIRS monitoring was
performed using the INVOS system (INVOS 5100 C;
Somanetics Corp., Troy, MI, USA). This system allows
non-invasive and continuous measurement of the
changes in the cerebral oxygen saturation. The NIRS
probes were placed in the frontotemporal region on
the eyebrows of the patients at both sides. The distance
between two probes was set as 4 cm. The rSO2 values
were recorded and certain parameters were evaluated
while the patients were awake before the induction
of anesthesia. After cross-clamps were placed, rSO2
changes of the patients during CPB were recorded in
the Excel format by the perfusionist and were evaluated
by two blinded observers. The standard algorithm was used,[13] as suggested for brain oximetry measurements,
in case of a >20% decline in the rSO2 d uring C PB,
compared to baseline.
Diffusion-weighted magnetic resonance imaging
Cranial d iffusion-weighted M RI w as p erformed
using the GE Signa HDxt MR device (GE Healthcare,
Waukesha, WI, USA). The patients were imaged in the
supine position using the standard GE 8-channel phased
array head-coil. Diffusion tensor imaging (DTI) was
measured using an echo planar imaging sequence with
TR/TE= 6000 ms/98.7 ms; 12 uniformly distributed
gradient directions, b-value of 1000 s/mm2, number
of excitations [NEX]= 2, contiguous axial slices with
5 mm slice thickness, 128¥128 matrix, and 27 cm
FOV. The slices were positioned to run parallel to
the anterior commissural and posterior commissural
plane. New diffusion-weighted lesions on postoperative
scans were also evaluated. Scans were read by a single
experienced neuroradiologist blinded to the clinical and
neuropsychological data of the patients until the time
of examination (before or after surgery). Preoperative
scans and new lesions are shown in Figures 1a-d.
New-generation oxygenators with integrated
arterial filters
The main difference between oxygenators
with integrated arterial filters and conventional oxygenators is represented by an embedded 30 micron
arterial filtering capability of the former technology
which avoids an increase in the total prime volume.
The maximum prime volume of the oxygenator is
260 mL including the hollow fiber membrane, heat
exchanger, and the arterial filter. The mid-upper part
of the oxygenator includes a bubble-trap, together
with a separate shunt line conveying the potentially
formed bubbles to the reservoir as well as a singledirection
valve at the oxygenator end of this line
which prevents backward flow. The pore size of the
venous/cardiotomy filter within the reservoir is less
than 26 microns. In addition, a 25-micron film is
incorporated into the cardiotomy reservoir to prevent
foaming.
Statistical analysis
Statistical analysis was performed using the IBM
SPSS version 23.0 software (IBM Corp., Armonk, NY,
USA). Descriptive analyses were performed to provide
data on overall characteristics of the study population.
The Kolmogorov-Smirnov test was used to test the
normality of the data. The independent sample t-test
was used to compare the patient characteristics of the
two groups (membrane oxygenators with or without integrated arterial filters). The paired sample t-test
was used to compare pre- and postoperative MoCA
scores. Two-way analysis of variance (ANOVA) for
repeated measurements was used to analyze the NIRS
values between two groups according to values of
cerebral saturation evaluated during surgery. The
continuous data were presented in mean ± standard
deviation. Categorical variables were compared using
the chi-square test and were presented in number
and percentage. A p value of <0.05 was considered
statistically significant.
Table 1: Demographic and clinical characteristics of patient groups
Table 2: Perioperative data, clinical outcomes and findings on brain magnetic resonance imaging
New MRI lesions were detected in 18 of 40 patients (45%), 11 in group 1 and seven in group 2 (p=0.34). The lesions identified in the postoperative diffusionweighted MRI were similarly distributed across the study groups (p=0.66). Of 18 patients with new lesions, four had a single lesion, four had two lesions, four had three lesions, and six had ≥3 lesions. The mean number of lesions was 2.8±7.4 and 1.9±3.8 in group 1 and group 2, respectively (p=0.631).
The mean preoperative total MoCA scores were 27.9±3.3 and 28.1±3.4 in group 1 and group 2, respectively (p=0.85). The mean postoperative total MoCA scores one month after surgery were 26.2±3.1 and 26.8±3.8 in group 1 and group 2, respectively (p=0.58). There was no statistically significant difference in the pre- and postoperative MoCA scores between the study groups (p=0.09 and p=0.26, respectively). In addition, a comparison of the change in MoCA test scores over time between the two groups did not show significant differences (p=0.11). The further analysis of MoCA scores are shown in Table 3.
Table 3: Mean Montreal Cognitive Assessment Scores according to the groups
In addition, the decline in saturation as assessed by the NIRS was similar between the two groups (p>0.05) and values are shown in Table 4.
Table 4: Mean near-infrared spectroscopy values according to the groups, time periods, and sides
Major cerebral injury represents a major cause of morbidity after cardiac surgery, occurring in up to 6% of patients undergoing CABG, as reported by recent studies.[13,14] Cognitive decline is a significant clinical marker of brain injury after cardiac surgery. Micro-embolism during CPB may also contribute to the postoperative cognitive decline (POCD) in these patients.[15] Recent evidence suggests that POCD may actually represent a multi-factorial condition.[15] However, regardless of the extent of the role of CPB in the development of POCD, it is recommended that every effort should be made to protect the patients from micro-embolism.[16] Newer generation oxygenators with integrated arterial filters have been introduced for the dual purpose of effective embolism filtration and improving CPB safety through reduced prime volume. The design of the newer oxygenators also allows incorporation of the arterial filter within the oxygenator chamber.[9,17] On the other hand, studies comparing conventional oxygenators without integrated arterial filters to those newer oxygenators with integrated arterial filters aiming at minimizing the risk of micro-embolism and POCD by forming less micro-bubbles and by entrapping more microbubbles are relatively low in number. To the best of our knowledge, the present study is the first prospective, randomized study to compare these two CPB oxygenator systems in terms of their effects on neurocognitive functions and neuroradiological outcomes. In the current study, the MoCA test was used to evaluate neurocognitive functions and cranial MRI was used to evaluate neuroradiological outcomes. To ensure the homogeneity of the cerebral oxygenation of patients in either group, the NIRS was also used for cerebral oxygenation monitoring. When needed, necessary interventions were performed to optimize the factors related to the cerebral brain oxygenation.
The NIRS is a tool, which is used to detect cerebral perfusion abnormalities, and cerebral MRI is used to detect cerebral lesions. In open heart surgery, the NIRS can offer invaluable information concerning the intraoperative cerebral perfusion alterations, which may be due to major bleeding and hemodynamic instability, during the aortic cross-clamping. Therefore, all patients in both groups were monitored by the NIRS through the operation in our study.
In a previous study by Gursu et al.,[9] Sorin Synthesis integrated arterial filtration system was compared with Sorin Dideco Compactflo (Dideco Compactflo Evo, Sorin group, Mirandola Modena, Italy) non-integrated arterial filtration system in patients undergoing CABG with respect to certain clinical variables, inflammatory responses, and need for transfusion. The authors did not report data on the incidence of embolism. However, they concluded that systems with integrated arterial filters were associated with higher intraoperative hematocrit levels and reduced need for red blood cell transfusions. In another study, oxygenators both with or without integrated arterial filters were found to convey micro-bubbles during the bolus entry of air into ECC and that bubbles larger than the filter pore size might be well-conveyed.[18] On the other hand, other authors reported reduced bubble volume and formation of smaller bubbles with integrated arterial filter systems than without it.[18] A recent study reported that the oxygenators without integrated filters exhibited the highest degree of emboli-trapping capabilities, although they found no disadvantage for oxygenators with integrated filters.[19]
The present study compared newly emerging embolic foci in cerebral diffusion-weighted MRI studies pre- and postoperatively in patients undergoing CABG using oxygenators with or without integrated arterial filters. Cerebral silent infarctions known to occur after cardiac procedures can be readily detected using diffusion-weighted MRI. Sun et al.[20] reviewed 12 recent reports in which diffusion-weighted MRI scans were used. Of the pooled population of 446 patients, 127 (29%) had new DWI lesions. Lesions were multiple and showed a widespread distribution across all cerebrovascular zones. Similarly, in our study, lesions were scattered across various cerebrovascular territories in both groups, showing no significant difference between the groups in terms of the distribution frequency as shown in Table 2. Also, in the aforementioned study, the occurrence of new lesions did not correlate with the type of cardiac surgery undertaken.[20] In a recent study, new diffusionweighted MRI lesions were found in five of 16 patients (31%) undergoing off-pump CABG.[21] In our study, new lesions were assessed using the same imaging modality and we found no statistically significant difference between the two groups in terms of the frequency and number of these lesions.
Nonetheless, there are certain limitations to our study. Our sample size was small, and although the study had a randomized prospective design, a similar study involving larger groups of patients would provide more meaningful results. Another limitation is the lack of a control group undergoing off-pump CABG. Also, only a single cognitive measurement tool was used for assessing the cognitive functions. This was based on the fact that the MoCA can be easily utilized in a busy clinical practice and enables health professionals to identify mild cognitive impairment through practical administration and interpretation steps.[5] Prolonged intensive care unit stay due to any reason, hemodynamic instability, and cerebrovascular event history were also among the exclusion criteria, which significantly limit the generalizability of our results to all patients undergoing cardiac surgery. In addition, ultrasonographic measurement of the gaseous micro-emboli at the entry of the venous site and way out of the arterial site might provide us more valuable results. Therefore, further large-scale studies including patients undergoing different types of cardiac surgery would provide more insight into the potential differences between these two oxygenator systems.
In conclusion, our results did not show any superiority of oxygenators with integrated arterial filters over those without integrated arterial filters in terms of cerebral outcomes based on assessments with cerebral diffusion-weighted imaging studies and cognitive tests. Therefore, further studies on larger series including control groups are required to corroborate our findings.
Declaration of conflicting interests
The authors declared no conflicts of interest with respect to
the authorship and/or publication of this article.
Funding
The authors received no financial support for the research
and/or authorship of this article.
1) Knipp SC, Matatko N, Wilhelm H, Schlamann M, Massoudy
P, Forsting M, et al. Evaluation of brain injury after
coronary artery bypass grafting. A prospective study using
neuropsychological assessment and diffusion-weighted
magnetic resonance imaging. Eur J Cardiothorac Surg
2004;25:791-800.
2) Roach GW, Kanchuger M, Mangano CM, Newman M,
Nussmeier N, Wolman R, et al. Adverse cerebral outcomes
after coronary bypass surgery. Multicenter Study of
Perioperative Ischemia Research Group and the Ischemia
Research and Education Foundation Investigators. N Engl J
Med 1996;335:1857-63.
3) Orhan G, Biçer Y, Aka SA, Sargin M, Simşek S, Senay
S, et al. Coronary artery bypass graft operations can be
performed safely in obese patients. Eur J Cardiothorac Surg
2004;25:212-7.
4) Abu-Omar Y, Balacumaraswami L, Pigott DW, Matthews PM,
Taggart DP. Solid and gaseous cerebral microembolization
during off-pump, on-pump, and open cardiac surgery
procedures. J Thorac Cardiovasc Surg 2004;127:1759-65.
5) Vohra HA, Whistance R, Modi A, Ohri SK. The inflammatory
response to miniaturised extracorporeal circulation: a review
of the literature. Mediators Inflamm 2009;2009:707042.
6) Liu Y, Tao L, Wang X, Cui H, Chen X, Ji B. Beneficial effects
of using a minimal extracorporeal circulation system during
coronary artery bypass grafting. Perfusion 2012;27:83-9.
7) Shann KG, Likosky DS, Murkin JM, Baker RA. An evidencebased
review of the practice of cardiopulmonary bypass
in adults: a focus on neurologic injury, glycemic control,
hemodilution, and the inflammatory response. J Thorac
Cardiovasc Surg 2006;132:283-90.
8) Riley JB. Arterial line filters ranked for gaseous microemboli
separation performance: an in vitro study. J Extra
Corpor Technol 2008;40:21-6.
9) Gürsu Ö, Isbir S, Ak K, Gerin F, Arsan S. Comparison of
new technology integrated and nonintegrated arterial filters
used in cardiopulmonary bypass surgery: a randomized,
prospective, and single blind study. Biomed Res Int
2013;2013:529087.
10) Katz ES, Tunick PA, Rusinek H, Ribakove G, Spencer
FC, Kronzon I. Protruding aortic atheromas predict
stroke in elderly patients undergoing cardiopulmonary
bypass: experience with intraoperative transesophageal
echocardiography. J Am Coll Cardiol 1992;20:70-7.
11) McLennan SN, Mathias JL, Brennan LC, Stewart S. Validity
of the montreal cognitive assessment (MoCA) as a screening
test for mild cognitive impairment (MCI) in a cardiovascular
population. J Geriatr Psychiatry Neurol 2011;24:33-8.
12) Nasreddine ZS, Phillips NA, Bédirian V, Charbonneau
S, Whitehead V, Collin I, et al. The Montreal Cognitive
Assessment, MoCA: a brief screening tool for mild cognitive
impairment. J Am Geriatr Soc 2005;53:695-9.
13) Kara I, Erkin A, Saclı H, Demirtas M, Percin B, Diler
MS, et al. The Effects of Near-Infrared Spectroscopy on
the Neurocognitive Functions in the Patients Undergoing
Coronary Artery Bypass Grafting with Asymptomatic
Carotid Artery Disease: A Randomized Prospective Study.
Ann Thorac Cardiovasc Surg 2015;21:544-50.
14) Wolman RL, Nussmeier NA, Aggarwal A, Kanchuger MS,
Roach GW, Newman MF, et al. Cerebral injury after cardiac
surgery: identification of a group at extraordinary risk.
Multicenter Study of Perioperative Ischemia Research Group
(McSPI) and the Ischemia Research Education Foundation
(IREF) Investigators. Stroke 1999;30:514-22.
15) Martin KK, Wigginton JB, Babikian VL, Pochay VE,
Crittenden MD, Rudolph JL. Intraoperative cerebral highintensity
transient signals and postoperative cognitive
function: a systematic review. Am J Surg 2009;197:55-63.
16) Johagen D, Appelblad M, Svenmarker S. Can the oxygenator
screen filter reduce gaseous microemboli? J Extra Corpor
Technol 2014;46:60-6.
17) Horton SB, Donath S, Thuys CA, Bennett MJ, Augustin SL,
Horton AM, et al. Integrated Oxygenator FX05. ASAIO J
2011;57:522-6.
18) Potger KC, McMillan D, Ambrose M. Air Transmission
Comparison of the Affinity Fusion Oxygenator with an
Integrated Arterial Filter to the Affinity NT Oxygenator
with a Separate Arterial Filter. J Extra Corpor Technol
2014;46:229-38.
19) Jabur GN, Sidhu K, Willcox TW, Mitchell SJ. Clinical
evaluation of emboli removal by integrated versus nonintegrated
arterial filters in new generation oxygenators.
Perfusion 2016;31:409-17.