Methods: A total of 31 patients (18 males, 13 females; mean age 51.5 years; range, 39 to 61 years) who were treated with long-term mechanical ventilation due to low cardiac output syndrome after cardiopulmonary bypass and cardiac surgery were retrospectively analyzed between December 2014 and December 2018. In addition, an in vitro lung model and different doses of hydroxyethyl starch in the heat and moisture exchangers to mimic the airway secretions were used and the proper interval to change heat and moisture exchangers was evaluated.
Results: In the in vitro l ung m odel, t he m ean a irway r esistance was 19.4±0.2 cmH2O/L/sec in the 5 mL group (p=0.060), 20.3±1.0 cmH2O/L/sec in the 10 mL group (p=0.065), and 30.2±1.7 cmH2O/L/sec in the 15 mL group (p<0.001). The airway resistance of heat and moisture exchangers, and total hospital stay and ventilation duration significantly increased in the seven-day group compared to the one-day and three-day groups. The positive culture of bacteria was also significantly higher in the seven-day group.
Conclusion: Our study results suggest that heat and moisture exchangers can be safely used for an efficient and timely removal of airway secretions. Volume of approximately 15 mL of liquid in the airflow can dramatically increase the airway resistance. The three-day interval of changing heat and moisture exchangers is ideal in a cardiothoracic surgery intensive care unit where patients have more airway secretions than patients in the general intensive care unit.
The HMEs increase airway resistance in patients who need long-term mechanical ventilation.[11] A number of studies have investigated whether HMEs can increase airway resistance in patients after CPB. Manufacturers recommend replacing disposable HMEs every 24 h;[12] however, Ricard et al.[13] showed that some HMEs could be changed once a week without any adverse effects for patients in general ICU. The proper intervals of changing HMEs in a cardiothoracic surgery ICU has not been well understood, yet.
In the present study, we aimed to investigate the efficacy and safety of HMEs in patients who underwent cardiac surgery and with a poor cardiac output, in the cardiothoracic surgery ICU. The first part of the study used an in vitro lung model fitted with an HME to mimic the effect of different amounts of airway secretions on airway resistance. The second part of the study retrospectively analyzed the efficacy and safety of changing HMEs at different intervals.
Figure 1: The experimental model of this study.
The lung model was ventilated using the ventilator. The ventilator settings were as follows: tidal volume (VT), 450 mL; respiratory rate, 13 bpm; positive end-expiratory pressure (PEEP), 5 cmH2O; and inspiratory time (IT), 2 s. A disposable HME and an ordinary disposable ventilator circuit were used. The ventilator circuit and lung model were checked before the experiment to ensure connection integrity and proper functionality. The HME was placed vertically above the tracheal tube (using a flex tube) to reduce the risk of partial obstruction due to refluxed secretions from the tracheal tube. The position of each HME was regularly checked. The HME was replaced before each measurement. The peak airway pressure (Ppeak) and platform pressure (Pplat) were monitored on the ventilator side of the HME during mechanical ventilation.[14]
The airway resistance was estimated based on the following equation:
Airway resistance=(Ppeak - Pplat)/V
The V was simulated based on the equation
V=VT/IT
during inspiration, and
Airway resistance=(Ppeak - Pplat)/(VT/IT)
V, inspiratory flow
The HMEs were connected to a nebulizer (Micro Mist® n ebulizer # 1884, Hudson RCI, CA, USA) that aerosolized settled dose of hydroxyethyl starch (HES) into the HME.[15,16] The amount of HES aerosolization was empirically determined to obtain a wet weight similar to the airway secretions that obtained by the patients after use. This weight information was derived from samples of each of the studied HMEs used in different patients after use.[11,17]
The HMEs were added 0 mL, 5 mL, 10 mL, and 15 mL of HES using a nebulizer. Four test groups (0 mL, 5 mL, 10 mL, and 15 mL) were formed, depending on the different doses of HES in HMEs, and one cohort without HME was used as control. The HMEs were connected in the ventilation circuit and the Ppeak and Pplat were monitored. The airway resistance can be estimated based on the monitored Ppeak, Pplat, VT, and IT.[14]
Retrospectively analysis of HME"s efficacy
and safety
Thirty-seven patients were retrospectively
analyzed in the department of cardiothoracic surgery
ICU between December 2014 and December 2018 at
the First Affiliated Hospital of Chongqing Medical
University. The patients were treated with long-term
mechanical ventilation due to low cardiac output
syndrome after CPB and cardiac surgery. Four
patients with pulmonary infections and two with other lung diseases, such as chronic obstructive
pulmonary disease (COPD), which affected the
duration of the mechanical ventilation and hospital
stay, were excluded. A total of 31 patients (18 males,
13 females; mean age 51.5 years; range, 39 to 61
years) were enrolled. A written informed consent
was obtained from each patient. The study protocol
was approved by the Ethics Committee of the First
Affiliated Hospital of Chongqing Medical University
(2014-033). The study was conducted in accordance
with the principles of the Declaration of Helsinki.
The patients were divided into three groups according to the interval of replacing HMEs. The HMEs for the first group were replaced daily, the second group every three days, and the third group every seven days. Humidity, bacterial culture of airway secretions, and airway resistance of HMEs were monitored. The total hospital stay and ventilation duration of each patient were also compared. The ventilator (Nellcor Puritan Bennett™ 840; Medtronic, CO, USA), ventilator circuit (Adult Breathing Circuit; Covidien LLC, Huizhou, Guangdong, China), and HME (PORTEX®, Smiths Medical International Ltd, Ashford, Kent, UK) were placed the same group as the in vitro lung model. The total follow-up time was seven days for each patient.
The number of tracheal suctioning and tracheal instillations was recorded daily. The Ppeak was monitored every 6 h and recorded as X±S. Absolute humidity (AH) and relative humidity (RH) were also measured. Absolute humidity at saturation (AHs) was used as a measurement of airway humidity, and it was the maximum amount of water vapor which could be contained in air at a given temperature. The RH was the ratio of AH-to-AHs, expressed in percentage. These parameters can be measured by psychrometry, a technique widely used in clinical studies to evaluate the performance of HME. The RH was calculated using a reference nomogram. The AH was calculated using the equation: AHs=16.451563-0.731T + 0.03987T2 mg H2O/L. T (ºC) is the temperature measured by a probe. The AH was derived using the formula: AH = (AHs × RH)/100 (in mg H2O/L).[18] The room temperature was kept constant at 24.0 to 26.0ºC.
The airway resistance of HMEs was measured daily and averaged according to the International Organization for Standardization (ISO)/Technical Committee (TC) 249 guidelines.[19] The airway resistance of HMEs in the one-day, three-day, and seven-day groups, when HMEs were replaced, was compared. Briefly, the airway resistance was calculated from the pressure drop on either side of the HME, when a constant 60-L/min flow was applied through the HME.[13]
To evaluate bacterial filterability of the HME, secretions drawn from the airway were sent for sputum culture, when HMEs were replaced during the study period.
Statistical analysis
Statistical analysis was performed using the IBM
SPSS version 21.0 statistical software (IBM Corp.,
Armonk, NY, USA). Descriptive data were expressed
in mean ± standard deviation (SD), median (min-max)
or number and frequency. Differences among the
groups were analyzed using the chi-square (χ2) test.
A p value of <0.05 was considered statistically
significant.
Table 1: Airway resistance in vitro lung model
Retrospectively analysis of HME"s efficacy
and safety
This study retrospectively studied 31 post-cardiac
surgery patients having low cardiac output and
mechanically ventilated for more than seven days.
This was to evaluate the efficacy and safety of
HME. Patients with pulmonary infections or other
lung diseases were excluded. Only synchronized
intermittent mandatory ventilation was used for
mechanical ventilation. Sex, age, smoking state, white
blood cell count, Acute Physiology and Chronic Health
Evaluation II (APACHE II) score, or partial pressure
of oxygen (PaO2)/fraction of inspired oxygen (FiO2)
were not significantly different in the use of HMEs
(Table 2).
Table 2: Demographic and clinical characteristics of patients
The parameters used to evaluate the efficacy and safety of HMEs are shown in Table 3. The number of tracheal inhalations, tracheal infusion volume, and the mean Ppeak were not significantly different among the one-day, three-day, and seven-day use of HMEs (p>0.05).
Table 3: Clinical assessment of humidifying efficacy
The AH and RH measured in the three-day and seven-day use of HMEs was not significantly different compared to changing HMEs daily throughout the study period (Table 4, p>0.05). Endotracheal tube obstruction never occurred during seven days of mechanical ventilation.
The airway resistance of HMEs was compared in one-day, three-day, and seven-day groups. The results indicate that airway resistance increased significantly in the seven-day group compared to the one-day group (p=0.002) and three-day groups (p=0.015, Table 5). There were no significant differences in the airway resistance of HMEs between the one-day and three-day groups (p=0.163). In addition, there was no significant difference in the ventilation duration and total hospital stay between the patients in the one-day and three-day groups (p>0.05). However, the total hospital stay was significantly longer in patients in whom the HMEs changed at a seven-day interval compared to the one-day interval group (p=0.030). The ventilation duration was significantly longer in patients in whom the HMEs changed at a seven-day interval compared to the one-day interval group (p=0.006). The total hospital stay and ventilation duration were longer in patients in the seven-day group compared to the three-day group (p=0.041 and p=0.021 respectively).
Table 5: Clinical assessment of airway resistance
The secretions were drawn from the airway and sent for bacterial culture. There were two positive cases of bacterial culture in one-day and three-day use of HMEs, and four positive cases in seven-day use of HMEs. The results showed that the seven-day group had a higher positive culture rate than the other groups (p<0.05).
The primary objective of this study was to evaluate airway resistance using the in vitro l ung m odel i n a ventilator fitted with an HME and to determine the quantity of secretions which would increase the airway resistance. The use of HMEs needs to prevent the sudden increase of airway resistance in patients on mechanical ventilation. The secondary objective was to compare the different intervals of changing HMEs (one-day, three-day, and seven-day intervals) by monitoring the humidity parameters, airway resistance, bacterial culture of the secretions, and duration of hospital stay and ventilation.
Airway resistance was higher in patients with airflow obstruction; however, Ikeda et al.[24] did not explore how much airflow obstruction led to a significant increase in the airway resistance. There are ISO standards for aerosolizing saline into the HME to evaluate the airway resistance, although the saline"s density and weight are different from the airway secretions of patients. In this study, we used HES to mimic actual airway secretions, and the density HES was similar to the airway sputum from the patients. Therefore, this study produced closer results to the real-life setting. Our results suggest that about 15 mL of liquid in the airflow can dramatically increase airway resistance. Patients tend to have more airway secretions in a cardiothoracic surgery ICU setting,[21] and the use of HME would increase the airway resistance.[14] Therefore, timely removal of airway secretions would be prudent during using HMEs.
Manufacturers recommend replacing disposable HMEs every 24 h, although objective data do not support this.[12] The weekly change o f HMEs was found to be effective and safe in general ICU through analysis of clinical parameters, humidity parameters, HMEs resistance of airflow, and bacterial colonization.[13] The proper interval of changing HMEs has not been well established in a cardiothoracic surgery ICU, due to the increased secretions after cardiothoracic surgery.
In our study, 31 patients on long-term mechanical ventilation were retrospectively analyzed due to low cardiac output syndrome after cardiac surgery (i.e., patients did not have signs of pulmonary infection or other lung diseases which may affect the airway resistance). These patients were divided into three groups based on the intervals of changing HMEs (one-day, three-day, and seven-day). Humidity, airway resistance of HMEs, secretion bacterial culture, total hospital stay, and ventilation duration were recorded. The results suggested that changing HMEs in a three-day interval did not show increased humidity, airway resistance, bacterial culture of secretions, total hospital stay, and ventilation duration of HMEs compared to daily changes. There was significantly increased airway resistance, bacterial culture, total hospital stay, and ventilation duration for HMEs changed at a seven-day interval, compared to daily change. This may due to the trends of patients to have longer surgical time, more trauma, and more airway secretions after cardiothoracic surgery than the general patients.[25] Jayakumar et al.[26] reported that cardiothoracic surgery had the most postoperative pain among surgeries. These factors can contribute to high airway secretions in patients after open chest surgery;[21] the changes in post-cardiothoracic surgery ICU patients with HMEs are different from general ICU patients. The three-day interval change of HMEs is ideal for patients with long-term mechanical ventilation undergoing cardiovascular surgery with CPB.
Airway secretions and sputum, which block the HMEs, may cause a sudden increase in airway resistance and eventually death.[27] Physicians and nurses in the ICU must inspect, if there are stains in the HMEs. The HMEs must be changed immediately, if stain occurs at any time.
This study has several limitations. First, the study represents one type of HME. Each type, design, and brand of HME is different for the amount of moisture collected;[25] therefore it is necessary to test a variety of HMEs to understand their effects on the airway resistance. Second, this study used HES instead of real airway secretions. Thus, the results may not exactly the same as the clinical situation, although the HES weight was the same as the airway secretions weight that obtained by the patients after use. Finally, further studies including larger samples are needed to confirm these findings.
In conclusion, our study results show that the safe use of heat and moisture exchangers facilitates efficient removal of airway secretions; only about 15 mL of liquid in the airflow can dramatically increase the airway resistance. It is not necessary to change heat and moisture exchangers daily. However, heat and moisture exchangers are needed to be replaced more than once a week in a cardiothoracic surgery intensive care unit and a three-day interval is ideal for patients on long-term mechanical ventilation. Patients who undergo cardiovascular surgery have more airway resistance than general intensive care unit patients. Changing heat and moisture exchanger at a three-day interval is safer than changed once a week and cheaper than daily.
Declaration of conflicting interests
The authors declared no conflicts of interest with respect to
the authorship and/or publication of this article.
Funding
This study was financially supported by the Science
and Technology Projects from the Science and Technology
Committee of Yuzhong district of Chongqing, (cstc2012ggyyjs10055)
China.
1) Aya HD, Cecconi M, Hamilton M, Rhodes A. Goal-directed
therapy in cardiac surgery: a systematic review and metaanalysis.
Br J Anaesth 2013;110:510-7.
2) Rhodes A, Moreno RP, Metnitz B, Hochrieser H, Bauer
P, Metnitz P. Epidemiology and outcome following postsurgical
admission to critical care. Intensive Care Med
2011;37:1466-72.
3) Pivatto Júnior F, Kalil RA, Costa AR, Pereira EM, Santos
EZ, Valle FH, et al. Morbimortality in octogenarian patients
submitted to coronary artery bypass graft surgery. Arq Bras
Cardiol 2010;95:41-6.
4) Gökalp O, İner H, Beşir Y, Gökalp G, Yılık L, Gürbüz
A. Using of Extracorporeal Membrane Oxygenation in
postcardiotomy heart failure. Turk Gogus Kalp Dama
2018;26:165-6.
5) Oruç M, Ülkü R. Evaluation of factors affecting prognosis
in penetrating thoracic injuries. Turk Gogus Kalp Dama
2018;26:598-605.
6) Tanigami H, Yahagi N, Kumon K, Watanabe Y, Haruna
M, Matsui J, et al. Long-term sedation with isoflurane in
postoperative intensive care in cardiac surgery. Artif Organs
1997;21:21-3.
7) Hummel J, Rücker G, Stiller B. Prophylactic levosimendan for
the prevention of low cardiac output syndrome and mortality
in paediatric patients undergoing surgery for congenital heart
disease. Cochrane Database Syst Rev 2017;8:CD011312.
8) Zarbock A, Mueller E, Netzer S, Gabriel A, Feindt P,
Kindgen-Milles D. Prophylactic nasal continuous positive
airway pressure following cardiac surgery protects from
postoperative pulmonary complications: a prospective,
randomized, controlled trial in 500 patients. Chest
2009;135:1252-9.
9) De Wet CJ, McConnell K, Jacobsohn E. Progress in
postoperative ICU management. Thorac Surg Clin
2005;15:159-80.
10) Martin C, Perrin G, Gevaudan MJ, Saux P, Gouin F. Heat
and moisture exchangers and vaporizing humidifiers in the
intensive care unit. Chest 1990;97:144-9.
11) Lucato JJ, Tucci MR, Schettino GP, Adams AB, Fu C, Forti
G Jr, et al. Evaluation of resistance in 8 different heat-andmoisture
exchangers: effects of saturation and flow rate/
profile. Respir Care 2005;50:636-43.
12) Kola A, Eckmanns T, Gastmeier P. Efficacy of heat and
moisture exchangers in preventing ventilator-associated
pneumonia: meta-analysis of randomized controlled trials.
Intensive Care Med 2005;31:5-11.
13) Ricard JD, Le Mière E, Markowicz P, Lasry S, Saumon
G, Djedaïni K, et al. Efficiency and safety of mechanical
ventilation with a heat and moisture exchanger
changed only once a week. Am J Respir Crit Care Med
2000;161:104-9.
14) Uchiyama A, Yoshida T, Yamanaka H, Fujino Y. Estimation
of tracheal pressure and imposed expiratory work of
breathing by the endotracheal tube, heat and moisture
exchanger, and ventilator during mechanical ventilation.
Respir Care 2013;58:1157-69.
15) van Galen G, Hallowell G. Hydroxyethyl starches in equine
medicine. J Vet Emerg Crit Care (San Antonio) 2019;29:349-59.
16) Ntoumenopoulos G, Berry MP, Camporota L, Lam L.
Indicators of Airway Secretion Weight in Mechanically
Ventilated Subjects. Respir Care 2019;64:1377-86.
17) Ahmed SM, Mahajan J, Nadeem A. Comparison of two
different types of heat and moisture exchangers in ventilated
patients. J Emerg Trauma Shock 2009;2:164-9.
18) Baskerville EB, Cobey S. Does influenza drive absolute
humidity? Proc Natl Acad Sci U S A 2017;114:E2270-E1.
19) Liu YQ, Wang YX, Shi NN, Han XJ, Lu AP. Current situation
of International Organization for Standardization/Technical
Committee 249 international standards of traditional Chinese
medicine. Chin J Integr Med 2017;23:376-80.
20) Hu XL, Xu ST, Wang XC, Hou DN, Chen CC, Song YL,
et al. Prevalence of and risk factors for presenting initial
respiratory symptoms in patients undergoing surgery for lung
cancer. J Cancer 2018;9:3515-21.
21) Wang Y, Li H, Zou H, Li Y. Analysis of Complaints
from Patients During Mechanical Ventilation After Cardiac
Surgery: A Retrospective Study. J Cardiothorac Vasc Anesth
2015;29:990-4.
22) Kartal E, Balaban ME. Machine learning techniques in cardiac
risk assessment. Turk Gogus Kalp Dama 2018;26:394-401.
23) Huang H, Yao T, Wang W, Zhu D, Zhang W, Chen H,
et al. Continuous ultrafiltration attenuates the pulmonary
injury that follows open heart surgery with cardiopulmonary
bypass. Ann Thorac Surg 2003;76:136-40.
24) Ikeda T, Uchida K, Yamauchi Y, Nagase T, Oba K,
Yamada Y. Relationship between pre-anesthetic and intraanesthetic
airway resistance in patients undergoing general
anesthesia: A prospective observational study. PLoS One
2017;12:e0172421.
25) Ari A, Dang T, Al Enazi FH, Alqahtani MM, Alkhathami
A, Qoutah R, et al. Effect of Heat Moisture Exchanger
on Aerosol Drug Delivery and Airway Resistance in
Simulated Ventilator-Dependent Adults Using Jet and
Mesh Nebulizers. J Aerosol Med Pulm Drug Deliv
2018;31:42-8.