Methods: Thirty patients who were multivessel completly revascularized with off-pump coronary artery bypass technique by using the octopus III stabilizator and Starfish heart positioner (Medtronic Minneapolis MN) were received in the study. The esophageal echo doppler system was used for monitorization of the hemodynamic parameters; cardiac output (CO), aortic blood flow (ABF), total systemic vascular resistance (TSVR), peak velocity (PV), stroke volume (SV), acceleration (Acc), left ventricular ejection time (LVET), heart rate, mean arterial blood pressure and aortic diameter.
Results: After clamping the left arterior descending artery (LAD) with bulldog clamp for anastomosis there was a significant decrease at the CO (Z = -2.584 p = 0.01). After reperfusing the LAD, CO increased the basic value in 2 minutes (Z = -0.773 p = 0.438) During PDA and Cx anastomosis there were no significant CO decreasing. After reperfusing all coronary arteries there was a significant increase (Z = 2.121 p = 0.034).
Conclusions: During the off-pump surgery the most important point is whether the heart tolaretes the heart position during off-pump surgery or not. Also monitorization technique has not been well accepted. The esophageal echo doppler offers accuracy equal that of the right heart catheter but in a noninvasive way.