ISSN : 1301-5680
e-ISSN : 2149-8156
Turkish Journal of Thoracic and Cardiovascular Surgery     
Resected bronchial blocker distal tip in bronchus stapler line: An unexpected complication
Tülay Hoşten1, Alparslan Kuş1, Salih Topçu2, Yavuz Gürkan1, Mine Solak1
1Departments of Anaesthesiology and Reanimation, Medical Faculty of Kocaeli University, Kocaeli, Turkey
2Departments of Thoracic Surgery, Medical Faculty of Kocaeli University, Kocaeli, Turkey
DOI : 10.5606/tgkdc.dergisi.2015.10758

Bronchial blockers (BBs) have become the preferred airway device in patients with particularly difficult airway, whose tracheobronchial anatomy is distorted, who may need postoperative mechanical ventilation, and are obese and children.[1] Therefore, complications associated with these airway devices should be well known.

A 62-year-old, 175 cm tall, 77 kg, American Society of Anesthesiologists physical status class 1 male patient was scheduled for left lower lobectomy. Following anesthesia induction, the patient was intubated with an 8.0 mm ID single lumen tube (SLT). A 9F, 78 cm Arndt bronchial blocker (BB) with elliptic balloon (Cook, Critical Care, Bloomington, USA) was placed in the left main bronchus under direct visual control with a 2.8 mm-diameter fiberoptic bronchoscope. Following right lateral decubitus positioning, the BB balloon was inflated and the optimal position was given (it closed the left main bronchus completely and the upper surface of the balloon did not exceed the main carina). During the bronchial incision step of the surgery, the balloon of the BB was deflated and removed, and the bronchus was closed with a stapler and cut off. When the bronchial staples line was checked at the surgical site, it was seen that the distal tip of the BB was stapled along with the bronchus at the stump line (Figure 1a, b). The staples were removed by the surgical team, the BB was pulled back and the bronchus was sutured with intermittent sutures, and the operation continued. At the end of the operation, the patient was extubated and no complications were observed.

Figure 1: (a) Cut tip of bronchial blocker in stump line. (b) Cut tip of Arndt bronchial blocker.

No major or life threatening complications were reported regarding independent BBs (Arndt, Cohen flexitip, Fuji Uniblocker, EZ blocker). As the manufacturer suggests in two case reports on Arndtrelated complications, the BB should be detached from the SLT along with its multiport adaptor and the wire in the central channel should be removed during one lung ventilation.[2,3]

The Arndt has outer diameter sizes of 5, 7, and 9 F and they are 50, 65 and 78 cm in length, respectively. Its balloon can be spherical or elliptical. When the elliptic-shaped Arndt BB balloon is inflated, it is 5 cm in length with the guide wire and 4 cm with the wire removed (Figure 2a, b). It must be remembered that the inflated balloon reaches as far as 3 cm deep into the bronchus after positioning the BB optimally. In the optimal position, the inflated end of the balloon is at about 0.5 cm distance from the carina.[4] Despite optimal positioning, this depth can be altered with manipulation during surgery. The anesthesiologist should know both the technical properties of BBs and tracheobronchial anatomy.[4] Before closing the surgical stapler, checking the site of BB by palpating the bronchus can be helpful in preventing this complication. After closing the bronchial stapler (prior to stapling), the retractability of the BB should be checked. Cooperation between the surgeon and the anesthesiologist bears extreme importance at this point.

Figure 2: (a) Arndt bronchial blocker with guide-wire. (b) Arndt bronchial blocker with inflated balloon and guide-wire removed during one-lung ventilation.

Preventing complications requires sound knowledge of tracheobronchial anatomy and technical properties of BBs, following suggestions of the manufacturer, and close cooperation between the surgeon and anesthesiologist.

Declaration of conflicting interests
The authors declared no conflicts of interest with respect to the authorship and/or publication of this article.

Funding
The authors received no financial support for the research and/or authorship of this article.