Elective TAVR with Medtronic-CoreValve®- Evolut-3R-26 system (Medtronic Inc. Minneapolis, Minnesota, USA) was performed in the catheterization laboratory under local anesthesia. In the final step, the CoreValve® prosthesis migrated to the ascending aorta (Figure 1, 2). As a result, an ECS was performed to retrieve the migrated prosthesis. Through median sternotomy, cardiopulmonary bypass was instituted with aortic cannulation at the aortic arch level to avoid the CoreValve® struts. The prosthesis was removed after filling the aortotomy site with cold saline to soften the rigid nitinol struts (Figure 3). Following complete resection of BAV and annular calcifications, a Carpentier-Edwards- Perimount-Magna-Ease-21 bioprosthesis (Edwards Lifesciences, Irwine, CA, USA) was implanted. Superior hemi-sternotomy was initially planned for the patient, if the decision of the Heart Team would go for SAVR. However, full-sternotomy was performed, due to ECS, and to provide full access the extraction of the prosthesis. The postoperative course was uneventful, and the patient was discharged postoperative 16th day.
The Heart Team discussion for assessing and informing the patient is unequivocal for an accomplished institutional TAVR program. Assessing the potential TAVR patients by the surgical team is important for two reasons: First, patients may be provided clear information concerning both treatment types (TAVR/SAVR) by each expert. Second, in case of peri-procedural complications, particularly for patients with a low surgical risk, ECS decision can be taken faster. Upper struts of the CoreValve® prosthesis in our case were seen under the adventitia of the ascending aorta without creating any local hematoma, dissection or bleeding. Quick decision of ECS, based on pre- TAVR consultation, may have prevented a fatal outcome, which might develop during an additional bail-out intervention.
In addition, TAVR should be performed in a hybrid room with the appropriate surgical armamentarium on stand-by, as recommended by current guidelines.[3,4] This is essential to gain time and increase the survival likelihood, in case of having major peri-procedural complications. As our case was hemodynamically stable following valve migration to the ascending aorta, we preferred to transfer her to the operating room to avoid postoperative infectious issues. In case of unstable or life-threatening conditions, the decision of performing the operation in the catheterization room should be balanced between infectious risks and mortality. Also, in a case series, Sener et al.[5] reported that the performance of TAVR procedures in catheterization rooms, where the sterilization guides are not as strict as they are in operating rooms, is a critical issue for prosthetic valve endocarditis following TAVR.
Although, the rate of major TAVR complications and mechanisms is controversial in the literature, the only common remark in all publications is the higher complication rates, including valve migration during the learning curve period.[6] In our case, based on the peri-procedural fluoroscopy images, we found that the mechanism of the CoreValve® prosthesis migration was incomplete disconnection from the delivery system of the new generation of CoreValve®, which has been recently adopted by our Cardiology Team. Therefore, a specifically trained Cardiology Team for these interventions should manage the patient, and during adoption of new techniques and TAVR valves, even if a former model was used, more vigilance and careful manipulations should be implemented.[7]
On the other hand, BAV is relative contraindication for TAVR, particularly for low surgical risk patients.[8] In registry reviews, patients with BAV who underwent TAVR had much higher risks of paravalvular leakage, malpositioning or migration, perhaps due to their elliptical annular anatomy, irregular (often severe) annular-leaflet calcifications, and possible less holding-wrapping resistance on the implanted valve than in tricuspid aortic valves.[9] Moderate-to-severe paravalvular leakage in BAV patients were more common than non-BAV patients.[9] Mylotte et al.[10] found the moderate-to-severe aortic regurgitation incidence as high as 28.4% in BAV patients, while it was only 9% in CoreValve® pivotal trial, and 12.2% in Partner 1A trial. The annular diameters or working mechanisms of the currently available catheter-based valves may not be always compatible with the calcified and deformed shape of the BAV. Balloon-expandable valves have lower regurgitation rates and, possibly, should be used more often than self-expandable valves to exclude such complications. In addition, Mylotte et al.[10] found almost half the prevalence of significant aortic regurgitation following with balloon expandable (19.6%) than with the selfexpandable (32.2%) valves in BAV patients.
In conclusion, this report supports the current concept that transcatheter aortic valve replacement is a promising option in high surgical risk patients, although it has still some limitations for low surgical risk patients, particularly those with bicuspid aortic valve. Appropriate selection of patients by the Heart Team and preoperative discussion of possible major complications and interventional plans may save patients life in case of lifethreatening complications. However, surgical aortic valve replacement remains standard of care for low surgical risk patients, particularly for those with bicuspid aortic valve.
Declaration of conflicting interests
The authors declared no conflicts of interest with respect to
the authorship and/or publication of this article.
Funding
The authors received no financial support for the research
and/or authorship of this article.
1) Loeser H, Wittersheim M, Puetz K, Friemann J, Buettner R,
Fries JW. Potential complications of transcatheter aortic valve
implantation (TAVI)-an autopsy perspective. Cardiovasc
Pathol 2013;22:319-23.
2) Eggebrecht H, Schmermund A, Kahlert P, Erbel R,
Voigtländer T, Mehta RH. Emergent cardiac surgery
during transcatheter aortic valve implantation (TAVI): a
weighted meta-analysis of 9,251 patients from 46 studies.
EuroIntervention 2013;8:1072-80.
3) Seiffert M, Conradi L, Baldus S, Schirmer J, Blankenberg S,
Reichenspurner H, et al. Severe intraprocedural complications
after transcatheter aortic valve implantation: calling for a heart
team approach. Eur J Cardiothorac Surg 2013;44:478-84.
4) Holmes DR Jr, Mack MJ, Kaul S, Agnihotri A, Alexander KP,
Bailey SR, et al. 2012 ACCF/AATS/SCAI/STS expert consensus
document on transcatheter aortic valve replacement: developed
in collabration with the American Heart Association, American
Society of Echocardiography, European Association for
Cardio-Thoracic Surgery, Heart Failure Society of America,
Mended Hearts, Society of Cardiovascular Anesthesiologists,
Society of Cardiovascular Computed Tomography, and Society
for Cardiovascular Magnetic Resonance. J Thorac Cardiovasc
Surg 2012;144:29-84.
5) Şener E, Küçüker A, Erdoğan K, Bayram H, Bilgiç A,
Durmaz T, et al. Prosthetic valve endocarditis after
transcatheter aortic valve implantation. Turk Gogus Kalp
Dama 2015;23:119-21.
6) Gurvitch R, Tay EL, Wijesinghe N, Ye J, Nietlispach F,
Wood DA, et al. Transcatheter aortic valve implantation:
lessons from the learning curve of the first 270 high-risk
patients. Catheter Cardiovasc Interv 2011;78:977-84.
7) Yılmaz O, Sönmez B, Ozturk S, Arbatlı H, Sener D, Baran T,
et al. Initial experiences with Corevalve® for transcatheter
aortic valve implantation in Turkey: two case reports. Turk
Gogus Kalp Dama 2013;21:769-72.
8) Colombo A, Latib A. Bicuspid aortic valve: any room for
TAVR? J Am Coll Cardiol. 2014;64:2340-2.