Methods: Between January 2012 and June 2015, a total of 73 patients (37 males, 36 females; mean age 71.8±5.7 years; range, 65 to 82 years) who underwent either surgical aortic valve replacement using supraannular bioprosthesis (n=36) or transcatheter aortic valve implantation (n=37) were included. Postoperative patient-prothesis mismatch was defined as absent, mild-to-moderate, and severe, if the indexed effective orifice area was >0.85 cm2/m2, >0.65 to <0.85 cm2/m2, and <0.65 cm2/m2, respectively. Both groups were compared in terms of patient-prothesis mismatch, postoperative outcomes, and mortality.
Results: The overall incidence of mild-to-moderate patient-prosthesis mismatch was 17.8% (13/73). No severe patient-prosthesis mismatch was observed. Mild-to-moderate patient-prosthesis mismatch was found in three patients (8.1%) in the transcatheter group and in 10 patients (27.8%) in the surgery group (p=0.035). Body surface area was the significant predictor of patient-prosthesis mismatch (p=0.007). Diameters of bioprosthetic valves in the surgery and transcatheter groups were 21.4±2 and 23.9±2.6 mm, respectively (p=0.002). Early mortality and pacemaker implantation rates were higher in the transcatheter group (p>0.05). Postoperative outcomes were similar between the groups. Mid-term mortality at a mean follow-up of 47.7±7.3 months was similar between the groups (p=0.158).
Conclusion: In high-risk patients with severe aortic stenosis, patientprosthesis mismatch is mild-to-moderate after surgical aortic valve replacement and transcatheter aortic valve implantation; however, this has no effect on early mortality. Based on our study results, we suggest that the use of surgical approach for aortic valve replacement may prevent potential complications of transcatheter aortic valve implantation.
Technical and technological innovations have provided new-generation valves for aortic position with improved EOA. The currently available transcatheter valves provide a better EOA, compared to surgical valves, as leaflets of these valves are mounted directly on the stents without a sewing ring.[5,10,11] On the other hand, annular calcifications are not resected during transcatheter procedures, in contrast to surgery, which may potentially affect EOA of transcatheter valves. Removal of annular calcifications during surgery and the use of a supra-annular bioprosthesis with a slim sewing ring above the aortic annulus can be associated with improved EOA after surgery. However, there is a limited number of data about the comparison of sAVR using a supra-annular bioprosthesis with TAVI in terms of PPM.[1,12] The use of surgical approach may also prevent potential complications of a TAVI procedure with unknown durability in long-term follow-up.
In the present study, we aimed to compare sAVR and TAVI in terms of development of postoperative PPM and to investigate the impact of PPM on postoperative outcome and mortality, as well as the procedure of first choice in high-risk patients.
Decision of sAVR or TAVI were obtained by the hospital council according to the international accepted guidelines.[12] Patients" characteristics, associated comorbidities, vascular access, expected functional outcome, and length of survival after procedures were considered in the selection of sAVR or TAVI in patients with a high surgical risk. All pre-, intra-, and postoperative data were collected from our surgical records and hospital medical network, and all the surgical notes and discharge summaries were reviewed to collect supplementary information. Demographic characteristics including age, gender, body mass index, body surface area (BSA), comorbidities such as hypertension, diabetes mellitus and obstructive lung disease, EuroSCORE index, ventricular functions, and valve features were recorded.
Preoperative work-up
Preoperative detailed transthoracic and
transesophageal echocardiography was performed to
all patients to evaluate the native valve, calcifications,
stenosis and insufficiencies, root and ascending
aorta. Coronary diseases were also eliminated
with preoperative angiography. Femoral artery
evaluation was performed via computed tomography
or angiography in patients who underwent TAVI
procedure.
sAVR technique
All AVR procedures were performed with standard
midline sternotomy and routine cardiopulmonary
bypass (CPB) by the same surgical team. In each
patient, the risk of PPM and the minimum size of the
prosthetic valve was evaluated preoperatively. After
resection of the aortic leaflets and decalcification
if needed, the surgeon measured the orifice area
carefully with an appropriate valve sizer, considered
body mass index of the patient, and replaced the native
valve with a suitable supra-annular bioprosthetic valve (Sorin Biomedica Cardio Srl, Sallugia, Italy).
The valves were implanted using separate 2/0 Ticron
sutures with pledgets. No root enlargement was
performed. All patients were weaned from CPB
uneventfully. The patients were followed in the
intensive care unit (ICU) postoperatively and, then,
transferred to the ward.
Transcatheter aortic valve implantation
technique
According to detailed evaluation of aortic annulus
and iliofemoral artery, transfemoral approach was
selected in TAVI procedures and the type and diameter
of the bioprosthetic valves were determined. The
type and number of the inserted bioprosthetic valves
were registered. The procedures were performed
under general anesthesia. After transfemoral access,
the procedures were done using either CoreValve
(Medtronic Inc., Minneapolis, MN, USA) or Edwards
Sapien (Edwards Lifesciences Ltd., Irvine, CA, USA)
bioprosthetic valves. Aortography and transesophageal
examinations were performed immediately to exclude
potential complications. Vascular access sites were
repaired using a closure device (Prostar XL, Abbott Vascular, Santa Clara, CA, USA) or surgically. The
patients were followed in the ICU postoperatively.
Postoperative follow-up
Postoperatively hospital and ICU stays, the
detailed echocardiographic features at three months
and follow-up period after surgery, early mortality
(within 30 days), and complications were recorded.
After discharge, the patients were followed regularly at
outpatient clinic under anticoagulant treatment.
Echocardiographic examinations
To determine the presence of PPM, the EOA of
each aortic valve surgically replaced were recorded
from their product guides. The EOA of each TAVI
valve was studied on postoperative transthoracic
echocardiography. Then, effective orifice area index
(EOAi) was calculated for every patient in postoperative
echocardiography. The presence and level of PPM was
recorded in every patient. The EOAi was found as
dividing EOA to BSA of the patient. Postoperative PPM
was defined as absent, mild-to-moderate, and severe, if
the EOAi was >0.85 cm2/m2, >0.65 to <0.85 cm2/m2,
and <0.65 cm2/m2, respectively.[1] Besides, the presence of paravalvular leak, aortic regurgitation, residual
gradient, and left ventricular function were studied.
Statistical analysis
Statistical analysis was performed using the
PASW for Windows version 17.0 software (SPSS
Inc., Chicago, IL, USA). Descriptive data were
expressed in mean ± standard deviation (SD) and
number and frequency. Other specific tests were
also used including variate analysis for the repetitive
measurements of multiple groups, the Student"s t-test
for the comparison of dual groups, chi-square test
for the comparison of qualitative data, and Pearson
correlation test to analyze the relationship between
the variables. A p value of less than 0.05 was
considered statistically significant.
Table 1: Preoperative demographic characteristics and echocardiographic findings
Preoperative echocardiography demonstrated that 20 patients (55.6%) had isolated and serious aortic valve stenosis and 14 patients (38.9%) had mixed-type aortic valve dysfunction in the sAVR group. Twentyfive (67.6%) of the TAVI patients had only aortic stenosis and mild aortic insufficiency accompanied with stenotic pathology in 12 patients (32.4%). Left ventricular EF in the sAVR and TAVI patients were 56.6±9.4% and 50.8±13%, respectively (p=0.031). The mean preoperative aortic valve gradients and aortic annulus diameters were similar between the groups (p=0.12 and p=0.51, respectively). The mean EuroSCORE values of the patients in the TAVI and sAVR groups were 8.9±3.8 and 7.3±7.0, respectively (p>0.05).
According to the calculation of the EOAi for each patient, severe PPM values (<0.65 cm2/m2) were not found in both either group (Table 2). However, in 17.8% (n=13) of all patients, mild-to-moderate PPM appeared. Three patients (8.1%) in the TAVI group and 10 patients (27.8%) in the sAVR group had mild-to-moderate PPM (p=0.035). The mean EOAi was 1±0.3 in the sAVR group and 1.2±0.2 in the TAVI group (p=0.004). The mean diameters of the inserted bioprosthetic valves in the sAVR and TAVI groups were 21.4±2 and 23.9±2.6, respectively (p=0.002).
Table 2: Presence of patient-prothesis mismatch according to effective orifice area index values
The mean postoperative length of ICU and hospital stays was 2.6±3.9 and 9.3±4 days in the sAVR group and 2.9±3.8 and 7.5±5.6 days in the TAVI group (p>0.05). Early mortality was defined as death within the first 30 days after the procedure or surgery. Of all patients, nine (12.3%) died; six (16.2%) in the TAVI group and three (8.3%) in the sAVR group (p=0.479) (Table 3). In the sAVR group, one of the patients died due to cerebrovascular event, and one due to postoperative low cardiac output syndrome. The remaining patient was admitted to the hospital with hemodynamic instability 20 days after discharge. Although echocardiography was normal at the time of discharge, serious paravalvular leak was detected in the second admission. This patient was the only one in whom moderate PPM was found among nonsurvivors. In the TAVI group, two patients died from cerebrovascular event, two from malignant arrhythmias and hemodynamic instability after the procedure. The remaining two patients died due to low cardiac output and multiorgan failure. The remaining two patients had higher EuroSCORE values as Log score= 21.18% and 18.76%, respectively.
Table 3: Postoperative data of patients
Postoperative data are presented in Table 3. There was no statistically significant difference between the groups in terms of the ratio of cerebrovascular event, pneumonia, arrhythmias and permanent pacemaker requirement, prolonged intubation, acute kidney failure, revision, and reoperation after the procedure or surgery. The mean postoperative gradients on the aortic valve were higher in the sAVR group than the TAVI groups (13.4±3.1 mmHg vs. 6.1±3.0 mmHg, p<0.05). When PPM-positive (+) and PPM-negative (-) patients were compared with each other, the BSA was found to be a significant predictor of PPM (p=0.007) (Tables 4 and 5). Mid-term mortality at a mean follow-up of 47.7±7.3 months was similar between the groups (TAVI vs. sAVR groups; 48.6% vs. 30.6%, respectively) (p=0.158) (Figure 1).
Table 5: Hospital and intensive care unit stays of PPM (+) and PPM (-) patients
Prosthesis-patient mismatch is a well-described event in patients undergoing aortic valve replacement procedures.[4,7,9-11] Age, gender, comorbidities, emergency of the operation, aortic root pathologies, infections, left ventricular EF, and selection of true size of prosthetic valve are the major factors that have close relation with mortality and PPM.[4-7,9] The concept of PPM was first introduced by Rahimtoola in 1978, as the effective prosthetic valve area, after insertion into the patient, is less than that of a normal human valve.[4-7] The main hemodynamic consequence of PPM is a high transvalvular pressure gradient through a normally functioning prosthetic valve after procedure. Therefore, the presence of PPM leads to reduced left ventricular mass regression in long-term and mortality.
Previous studies showed that postoperative PPM might have a significant impact on cardiac events and mortality in both sAVR and TAVI procedures.[4-11] However, in our study, we found no statistically significant difference between the PPM (+) and PPM (-) groups including early mortality and complications. In addition, no severe PPM was found in any patients, which could be considered an acceptable outcome of both sAVR and TAVI procedures. This finding indicates that open surgery is still the superior approach in high-risk patients, despite higher EOA and EOAi of TAVI valves. Although mild-to-moderate PPM was higher in the sAVR patients, postoperative outcomes and morbidities were similar between the groups. However, the incidence of pacemaker implantation and early mortality was higher in the TAVI group. Despite relatively short follow-up period, we may conclude that the presence of PPM has no impact on the early postoperative events.
There is a limited number of studies comparing sAVR and TAVI in terms of postoperative outcomes, mortality, and PPM. Cohort A section of the PARTNER trial showed 30-day mortality rates for AVR and TAVI as 6.5% and 3.4% respectively. However, oneyear mortality rates were found to be 26.8% for sAVR and 24.3% for TAVI.[1,14,17,18] On the other hand, postoperative complications, cerebrovascular events, paravalvular leakage, and functional aortic valve insufficiency were found to be more frequent for TAVI in the aforementioned trial. In our study, however, we found no statistically significant difference in the TAVI and sAVR groups in terms of early mortality and postoperative complications. Moreover, regarding postoperative PPM, the PARTNER trial showed that the incidence of PPM was 60.0% (severe: 28.1%) in the sAVR-randomized control trial cohort versus 46.4% (severe: 19.7%) in the transcatheter aortic valve replacement (TAVR)-randomized control trial cohort (p<0.001) and 43.8% (severe: 13.6%) in the TAVR-nonrandomized control trial cohort.[1] This study is the latest and the only report regarding the comparison of surgery and TAVI for PPM. The authors concluded that, in patients with severe aortic stenosis and high surgical risk, mild-to-moderate PPM was more frequent after sAVR than TAVI. Of note, the absence of severe PPM should be considered an advantage of supra-annular bioprosthetic aortic valves.
Several risk factors including female gender, obesity, age >65 years, and New York Heart Association (NYHA) Class III/IV heart failure have been reported to be the major predictors of clinically significant PPM.[4-9] Patients w ith aortic stenosis tend to have a smaller aortic annulus, which not only allows for a smaller valve to be placed, but also decreases the postoperative EOA; this, in turn, increases the risk of postoperative PPM. In the PARTNER trial, the results showed that, in patients with an aortic annulus diameter <20 mm, severe PPM developed in 33.7% undergoing sAVR compared to 19.0% undergoing TAVR (p<0.002).[1] The authors concluded that TAVI might be preferable to sAVR in patients with small aortic annulus. In our study, PPM (+) and PPM (-) patients were also compared with each other and showed similarity regarding preoperative demographic features, postoperative mortality, and ICU and hospital stay. However, we observed that the only predictor of PPM was low BSA, which was another risk for PPM, even if the sizes of implanted valve were similar.
In the present study, we detected 13 (17.8%) mild-moderate PPM (+) patients. No severe PPM was observed in all patients. Two groups were compared with each other regarding PPM and PPM was found more frequent (n=10) in sAVR group significantly (p=0.035). Patients had similar demographic features and EuroSCORE values in two groups. Although studies are limited in the literature comparing TAVI with sAVR in terms of PPM, TAVI has often better results in recent studies.[1,10,11] In the PARTNER trial, the mean gradient ratio and EOA ratio were found better in the TAVI.[14] Therefore, it is necessary to evaluate the increased rates of PPM in surgery. A more detailed evaluation of patients using preoperative echocardiographic measurements and excluding patients with small aortic root and orifice may lead to better results with surgical intervention. Some technical options such as aortic root enlargement, selection of true size of valve, selection of supra-annular valves, or stentless bioprosthesis and management of difficulties during implantation of the bioprosthesis may prevent PPM after surgery.
In our study, mortality of patients in the TAVI group was higher than patients in the sAVR group; however, we found no statistically significant difference between the groups. The results of the PARTNER 1 trial showed that, at five years, the risk of death was 67.8% in the TAVR group compared to 62.4% in the sAVR group (p=0.76).[19] In addition, PPM, particularly in the elderly group, does not increase the mortality in many of the studies. Thus, we may conclude that reducing the PPM may be non-significant in this elderly population. Similarly, the presence of mild-to-moderate PPM following bioprosthetic sAVR does not influence long-term survival.[12,14] With the unknown durability of TAVI valves, the increased need for permanent pacemakers in these valves which has been shown to be associated with decreased survival, and the lack of follow-up data, we may not conclude that TAVI may be the initial choice of treatment in patients with an increased BSA and small aortic annulus, who are at risk of PPM. The use of sAVR may prevent potential complications of a TAVI procedure and implantation of a TAVI valve with the unknown durability in long-term.
Although prostheses have the same number and geometric orifice areas, functional EOA can be different depending on the BSA, model of the prothesis and gender of the patient. Different pressure gradients develop in different patients, since determinative phenomenon of the cardiac output at rest is BSA. The EOAi depending on the BSA of the patient has 80% specificity and 74% sensitivity to estimate the presence of PPM.[4] After all, it has been noted that PPM can be largely avoided using the correct preventive strategy at the time of operation both for sAVR and TAVI.[1-18] In surgery, the incidence of PPM can be reduced by choosing the correct size of bioprosthesis, using stentless bioprosthesis, supra-annular implantation of the valve, aortic root manipulations or extensions, considering the left ventricular EF, BSA, age, and gender of the patient during operation. Recent studies have also demonstrated that allografts, homografts, and stentless bioprostheses have less gradient differences inversely greater EOA. In the coming days, these grafts can be used more frequently both in surgery and TAVI.[1-3,19,20]
Nonetheless, there are some limitations to this study. Small sample size in both groups and relatively short follow-up period to examine long-term effects of PPM on cardiac events are the main limitations. Retrospective design of the study is also another limitation. In addition, the lack of the aortic root enlargement procedures during operations can be considered a limitation in the surgical patient group. Additionally, the mean EF of patients with PPM was 58.8±5.8% which was statistically significant. This might have brought about a bias which is that the PPM might be due to high EF. As EF may lead to miscalculation of the transvalvular gradients, this bias is a limitation of the study.
In conclusion, in high-risk patients with severe aortic stenosis, patient-prosthesis mismatch is mild-tomoderate after surgical aortic valve replacement and transcatheter aortic valve implantation; however, this has no effect on early mortality. With the unknown durability of transcatheter aortic valve implantation valves, the increased need for permanent pacemakers in these valves, which has been associated with decreased survival, and the lack of long-term follow-up data, we conclude that the use of surgical aortic valve replacement may prevent the potential complications of transcatheter aortic valve implantation. Nevertheless, surgeons should perform aortic valve replacement with root enlargement techniques to decrease the rate of postoperative patient-prosthesis mismatch.
Declaration of conflicting interests
The authors declared no conflicts of interest with respect to
the authorship and/or publication of this article.
Funding
The authors received no financial support for the research
and/or authorship of this article.
1) Pibarot P, Weissman NJ, Stewart WJ, Hahn RT, Lindman BR,
McAndrew T, et al. Incidence and sequelae of prosthesispatient
mismatch in transcatheter versus surgical valve
replacement in high-risk patients with severe aortic stenosis:
a PARTNER trial cohort--a analysis. J Am Coll Cardiol
2014;64:1323-34.
2) Gedik HS, Korkmaz K, Lafçı G, Yalçınkaya A, Çağlı K. The
effects of implanted valve sizes on ventricular hypertrophy in
aortic stenosis. [Article in Turkish] Anadolu Kardiyol Derg
2012;12:165-70.
3) Uluçam M. Diagnosis and treatment of prosthetic valve
dysfunction. [Article in Turkish] Anadolu Kardiyol Derg
2009;9:35-42.
4) Pibarot P, Dumesnil JG, Cartier PC, Métras J, Lemieux MD.
Patient-prosthesis mismatch can be predicted at the time of
operation. Ann Thorac Surg 2001;71:265-8.
5) Clavel MA, Webb JG, Pibarot P, Altwegg L, Dumont
E, Thompson C, et al. Comparison of the hemodynamic
performance of percutaneous and surgical bioprostheses for
the treatment of severe aortic stenosis. J Am Coll Cardiol
2009;53:1883-91.
6) Kukucka M, Pasic M, Dreysse S, Mladenow A, Habazettl H,
Hetzer R, et al. Patient-prosthesis mismatch after transapical
aortic valve implantation: incidence and impact on survival.
J Thorac Cardiovasc Surg 2013;145:391-7.
7) Bakir I, Casselman FP, Onan B, Van Praet F, Vermeulen Y,
Degrieck I. Does a minimally invasive approach increase
the incidence of patient-prosthesis mismatch in aortic valve
replacement? J Heart Valve Dis 2014;23:161-7.
8) Dumesnil JG, Pibarot P. Prosthesis-patient mismatch: an
update. Curr Cardiol Rep 2011;13:250-7.
9) Head SJ, Mokhles MM, Osnabrugge RL, Pibarot P,
Mack MJ, Takkenberg JJ, et al. The impact of prosthesispatient
mismatch on long-term survival after aortic valve
replacement: a systematic review and meta-analysis of 34
observational studies comprising 27 186 patients with 133
141 patient-years. Eur Heart J 2012;33:1518-29.
10) Smith CR, Leon MB, Mack MJ, Miller DC, Moses JW,
Svensson LG, et al. Transcatheter versus surgical aorticvalve
replacement in high-risk patients. N Engl J Med
2011;364:2187-98.
11) Kodali SK, Williams MR, Smith CR, Svensson LG, Webb
JG, Makkar RR, et al. Two-year outcomes after transcatheter
or surgical aortic-valve replacement. N Engl J Med
2012;366:1686-95.
12) Nishimura RA, Otto CM, Bonow RO, Carabello BA, Erwin
JP 3rd, Guyton RA, et al. 2014 AHA/ACC guideline for
the management of patients with valvular heart disease:
executive summary: a report of the American College of
Cardiology/American Heart Association Task Force on
Practice Guidelines. J Am Coll Cardiol 2014;63:2438-88.
13) Brown JM, O'Brien SM, Wu C, Sikora JA, Griffith BP,
Gammie JS. Isolated aortic valve replacement in North
America comprising 108,687 patients in 10 years: changes in
risks, valve types, and outcomes in the Society of Thoracic
Surgeons National Database. J Thorac Cardiovasc Surg
2009;137:82-90.
14) Lefèvre T, Kappetein AP, Wolner E, Nataf P, Thomas M,
Schächinger V, et al. One year follow-up of the multi-centre
European PARTNER transcatheter heart valve study. Eur
Heart J 2011;32:148-57.
15) Gilard M, Eltchaninoff H, Donzeau-Gouge P, Chevreul K,
Fajadet J, Leprince P, et al. Late Outcomes of Transcatheter
Aortic Valve Replacement in High-Risk Patients: The
FRANCE-2 Registry. J Am Coll Cardiol 2016;68:1637-47.
16) Généreux P, Head SJ, Wood DA, Kodali SK, Williams MR,
Paradis JM, et al. Transcatheter aortic valve implantation
10-year anniversary: review of current evidence and clinical
implications. Eur Heart J 2012;33:2388-98.
17) Lawrie GM. Role of transcatheter aortic valve implantation
(TAVI) versus conventional aortic valve replacement in
the treatment of aortic valve disease. Methodist Debakey
Cardiovasc J 2012;8:4-8.
18) Leon MB, Smith CR, Mack M, Miller DC, Moses JW,
Svensson LG, et al. Transcatheter aortic-valve implantation
for aortic stenosis in patients who cannot undergo surgery. N
Engl J Med 2010;363:1597-607.