Figure 2: Two-cm metallic part of endostapling device that was missing (circled).
Explorative visualization as well as counting surgical instruments and gauzes are routinely performed before closure in each VATS as a surgical principle. Methodologies including Six Sigma that strive to achieve perfect quality[8] may be adopted as a quality principle for surgeries to prevent such structural problems.[8,9]
This case should remind us about the importance of double-checking the integrity of the used stapling devices during surgery. All team members in the operating room including the circulating nurse should be aware of the necessity to check such details to prevent any event that may require an additional surgery.
Declaration of conflicting interests
The authors declared no conflicts of interest with respect to
the authorship and/or publication of this article.
Funding
The authors received no financial support for the research
and/or authorship of this article.
1) Onaitis MW, Petersen RP, Balderson SS, Toloza E, Burfeind
WR, Harpole DH Jr, et al. Thoracoscopic lobectomy is a safe
and versatile procedure: experience with 500 consecutive
patients. Ann Surg 2006;244:420-5.
2) Whitson BA, Groth SS, Duval SJ, Swanson SJ, Maddaus
MA. Surgery for early-stage non-small cell lung cancer:
a systematic review of the video-assisted thoracoscopic
surgery versus thoracotomy approaches to lobectomy. Ann
Thorac Surg 2008;86:2008-16.
3) Amosov NM, Berezovsky KK. Pulmonary resection with
mechanical suture. J Thorac Cardiovasc Surg 1961;41:325-35.
4) Weissberg D, Kaufman M. Suture closure versus stapling
of bronchial stump in 304 lung cancer operations. Scand J
Thorac Cardiovasc Surg 1992;26:125-7.
5) Yano M, Iwata H, Hashizume M, Shigemitsu K, Adachi
K, Tokui T, et al. Adverse events of lung tissue stapling
in thoracic surgery. Ann Thorac Cardiovasc Surg
2014;20:370-7.
6) Yano M, Takao M, Fujinaga T, Arimura T, Fukai I, Ota S, et
al. Adverse events of pulmonary vascular stapling in thoracic
surgery. Interact Cardiovasc Thorac Surg 2013;17:280-4.
7) Gossot D, Merlusca G, Tudor A, Boudaya MS, Radu C,
Magdeleinat P. Pitfalls related to the use of endostaplers
during video-assisted thoracic surgery. Surg Endosc
2009;23:189-92.