A novel device for technical standardization of valve-sparing aortic root reimplantation
Ahmet Elibol1
1Department of Cardiovascular Surgery, University of Health Sciences, Kartal Koşuyolu Yüksek İhtisas Training and Research Hospital, Istanbul, Turkey
DOI : 10.5606/tgkdc.dergisi.2020.19182

Abstract

Among valve-sparing aortic root replacement techniques developed for the treatment of aortic root aneurysms and aortic insufficiency, the reimplantation technique (the David procedure) has proved to provide excellent outcomes in experienced hands. However, it involves certain challenges in technical standardization, particularly for graft sizing, which is still far from standardization. A novel device was developed to facilitate and provide all the measurements in high precision and accuracy required for the David procedure. The device allows easy, rapid, and accurate acquisition of the patient"s data and appropriate configuration of the aortic valve, irrespective of the surgeon's subjective evaluations. This all-in-one device provides all the major parameters including graft size, effective height, graft preparation, and simulation of the aortic coaptation. The device was successfully tested on a Devotini aortic root simulator and on a bovine heart ex vivo. The device proposed herein to be used for reimplantation has one explicit advantage: all valve geometry to be reconstructed and repaired can be simulated on the device with all its elements, in particular, the commissures and the cusps. Thus, all that is necessary can be clearly visualized in a manner whatever the configuration the surgeon prefers, particularly the creation of the effective height.

Aortic root replacement surgery is performed in patients with aortic root aneurysms. Besides the threat of dissection, aortic root aneurysms may also cause aortic valve regurgitation due to dilatation of the base of the aortic valve (Figure 1). The most common practice is to perform the Bentall procedure, in which the aortic root aneurysm is removed along with the aortic valve. In this procedure, the native aortic valve is replaced with a mechanical or biologic prosthetic valve, even if the leaflets are not involved. However, both types of prosthetic valves are associated with severe complications.[1-3] Recipients of mechanical aortic valves are at risk for infections and thromboembolism, for which life-long anticoagulant therapy is indicated, with regular hospital visits and dose adjustment. In addition, mechanical aortic valves impose significant limitations on the patients" lifestyle, restricting activity and for women, pregnancy. Although biologic prosthetic valves do not require anticoagulant therapy, structural deterioration is inevitable over time[4] and they have been shown to be associated with higher mortality and reoperation rates before 55 years of age.[3]

Figure 1: Anatomy of aortic root aneurysms.

Due to these limitations, valve-sparing aortic root replacement techniques have been developed to preserve the native aortic root, including reimplantation and remodeling techniques.[5,6] The reimplantation technique (David procedure) has been progressively gaining interest owing to the wide range of indications it offers.[7] In this procedure, the aortic annulus is dissected and reimplanted into the replacement graft. However, it requires patient-specific anatomical measurements of the aortic root and valve, such as appropriate graft size, commissure height (height of the interleaflet triangle), and effective height, which are difficult to estimate and require experience.[5] Several techniques (e.g., echocardiography, computed tomography [CT], transesophageal echocardiography [TEE], valve sizers, or Hegar dilators) have been used to identify the appropriate graft size.[8] Commissure height is also used to estimate the graft size; however, this technique is highly based on visual inspection. Effective height, on the other hand, is determined using an aortic caliper, but accuracy depends on its vertical positioning on the aortoventricular plane. Therefore, the success of the David procedure highly depends on the skills and experience of the cardiothoracic surgeon. Thus, rather than leaving all these measurements to the surgeon"s discretion and preference, which is highly dependent on the surgeon"s skills and experience, there is an unmet need to develop a device and method to make standard and accurate measurements of the anatomical features of the aortic root and leaflet geometry.[9]

In recent years, a novel device has been developed to facilitate and provide all the measurements in high precision and accuracy required for the David procedure, specifically for David type I reimplantation. The device was patented to the developer (author AE) by No. TR/2016/18240 B. The device has three main advantages: (i) it provides the appropriate graft size based on an objective determination of the effective height; (ii) all measurements required for the David procedure can be obtained on the device, free from individual interpretations; and (iii) the device provides excellent guidance to graft preparation convenient for the patient"s anatomy and to placement of the graft in position.

THE NEED FOR A NEW DEVICE FOR THE DAVID PROCEDURE
THE MAJOR ELEMENTS OF THE DAVID PROCEDURE

Valve-sparing aortic root reimplantation (VSARR) surgery requires an accurate measurement of the aortic valve and the root. The key to success for VSARR is the choice of appropriate graft size, which depends on the accurate measurement of the patient"s anatomy. Miller DC, a distinguished cardiovascular surgeon with expertise in the David procedure, rightly commented on the procedure as follows: "This procedure remains more art than science and is most unforgiving of small technical errors in dynamic three-dimensional geometry."[10]

Graft sizing
The sizing of the graft, appropriate in diameter for the patient, is of utmost importance to achieve an optimal coaptation of the aortic leaflets and to avoid postoperative aortic insufficiency.[11] Several methods have been described for the selection and preparation of the graft.[12,13] Therefore, this stage is still far from standardization and remains mainly at the discretion of the operating surgeon. A graft of smaller than optimal size results in prolapse (collapse) with ensuing aortic regurgitation, while selection of a larger graft results in central aortic regurgitation.

The first step in choosing the right size of the graft is the measurement of the aortic annulus. Several methods have been described for this measurement, most of which are based on the measurement of the aortic annulus including pre- and intraoperative TEE, preoperative CT, sizing with Hegar rods to calculate the inner diameter of the aortic annulus, and sizing with valve sizers, where a mechanical valve sizer is fitted around the annulus and a biological valve sizer is placed inside the aortic annulus.[11] If the selected method measures the inner diameter of the aortic annulus, the graft size is, then, often determined by adding 4 mm to the diameter.[11] However, D avid h as objections to the use of anatomic measurements of the annulus made based on the assumption that the aortic annulus is already normal, since the anatomic components of the aortic valve are always disrupted due to aortic root aneurysms.[14] Due to the complexity of the measurements and the geometric variables of the aortic apparatus, David[14] proposed an indirect means of determining the graft size and estimated the graft size in relation to the average height of the cusps. Given that the radius of the reconstructed aortic root should be approximately two-thirds of the average height of the cusps, an addition of 6 mm is made to the two-thirds of the average height of the aortic cusps due to the placement of the graft outside of the aortic annulus. On the other hand, El Khoury[15] proposed the commissural height lying between the non-coronary cusp and the left coronary cusp to be used for determining the graft size, with the height of the commissure being equal to the external diameter of the sinotubular junction, equivalent to the graft diameter.

It is evident that the proposed measurement methods for the graft size are quite confounding and far from standardization.[16] The choice of the most optimal technique and the success of VSARR, therefore, often depends on the experience and individual preference of the surgeon.

Effective height
After reimplantation of the aortic valve inside the graft, it is important that all three cusps coapt at the same level, which should be at least 5 or 6 mm above the aortic annulus according to David. If the reimplantation does not provide coaptation at the same level, shortening of the free margin that disrupt coaptation may be needed.[14]

On the other hand, Schäfers and Aicher[17] introduced the use of an aortic caliper to measure the effective height and to check for the appropriate coaptation so that any residual or induced prolapse can be repaired. The use of an aortic caliper has been advocated by other cardiovascular surgeons, as well.[18] The recommended range for the effective height is 7 to 11 mm, depending on individual surgeons.[8,19]

A study which examined the effective height by transthoracic echocardiography in healthy adults found a mean effective height of 9.5 mm (range, 7 to 12) with a significant correlation with the body surface area, body weight, and body height.[20]

Subannular suture line and graft preparation
The placement of subannular sutures to fix the graft below the annulus has a significant role on leaflet configuration, coaptation, and effective height. The arrangement of these sutures is highly patient-specific, although there are variations in the preparation of the graft in relation to the site of the subannular sutures. As the subannular suture line does not lie in the same plane, the placement of the graft on the aortic root must match the alignment of the subannular suture line so that the corresponding points of the suture line can be determined on the graft.[21]

Reimplantation of the aortic commissures inside the graft
All three commissures should be positioned and reimplanted properly in the graft to form the new sinotubular junction, paying an attention to their spatial relationships with each other.[5] This part of the operation requires knowledge and experience to reimplant the valve exactly in the same position as it was before in the native aortic root.[11] If the commissures are fixed too low or high, this may lead to aortic insufficiency due to leaflet prolapse and oversuspension, respectively.

INNOVATION
The novel device, Elibol"s David facilitator, measures the above-mentioned parameters, independent of the surgeon's subjective assessments, for the success of the David procedure. It also simulates coaptation at the very beginning of the procedure.

Once the device is seated on the aortic root, the surgeon will be able to obtain all data about these major parameters within only approximately 10 min and can proceed with the David procedure without having to choose from a wide variety of currently used methods or techniques.

Design of the device
The main components of the device include (i) an adjustable circle for graft diameter (26 to 34 mm), (ii) f ive l egs f or f ixation o utside t he a ortic a nnulus, each having a spring, (iii) three commissure holders attached to the circle, and (iv) two pins (central and peripheral) that serve as calipers to perform effective height measurement (Figure 2). The adjustable circle moves vertically on the legs to change the height of the commissure holders and horizontally (narrowing and widening) to change the distance of the commissure holders from each other. The legs are fixed to the aortic annulus at the subannular suture line using slit surgical tubes. The central and peripheral calipers measure the vertical heights from the coaptation tip and the nadir of each leaflet, respectively. The difference yields the effective height. The components and a step-by-step guide to the application of the device are provided in the supplementary material.

Figure 2: The appearance of device and its components.

Functions of the device
The all-in-one Elibol's David facilitator was designed to eliminate the variability associated with diverse measurement techniques. (i) The device simulates optimum leaflet coaptation by modifiable resuspension through commissure holders on the adjustable circle. (ii) Once t he optimal coaptation is achieved, the effective height for each leaflet (as the subtraction of the height of the coaptation tip from the height obtained at the nadir of each leaflet) is checked. At this stage and before graft selection, aortic cusp repair may be performed and/or effective height can be modified for individual leaflets in case optimal coaptation cannot be obtained. (iii) The device is locked and removed from the aortic annulus, at which time all measurements can be read on the device including the graft size as defined by the diameter of the circumference determined by the commissure holders. (iv) The graft chosen at the same diameter is inserted inside the legs of the device and attached to the commissure holders. (v) The subannular suture line and the resuspension points of the commissures are readily marked on the graft. (vi) The graft is cut just b elow the suture line. (vii) The r emaining p art of the reimplantation is merely tailoring.

EXPERIMENTAL STUDIES
Testing the device on an aortic root simulator

A three-leg prototype of the device was tested on a Devotini aortic root simulator. The ascending aorta and the sinuses of Valsalva were resected (Figure 3a). Subannular sutures that would form the subannular suture line were placed as in the David procedure, and the three aortic commissures were created with varying commissural heights in relation to the subannular suture line (Figure 3b). The device was, then, placed in the aortic root under the guidance of subannular sutures that corresponded to the three legs (Figure 3c). The commissure holders were attached to the three aortic commissures, above the commissural insertions of the leaflets (Figure 3d). While the springs on the legs pushed the legs downwards and the circle upwards, the commissure holders attached to the commissures limited the upward movement of the circle. This provided vertical positioning of the device on its own mechanics, which was observed to be stable (Video 1), in that its position returned to the initial state, even if an external force was applied to change the position. This fixed position provides a steady plane above the commissures (the new sinotubular junction), keeping the adjustable circle parallel to the subannular plane at each position of the heart. Simulation of coaptation was observed at varying diameters of the circle (Figure 4a, b). Once an optimal coaptation was observed (Figure 4c), the legs were locked and the effective height was obtained using the two (central and peripheral) pins (Figure 5a, Video 2). The device was, then, removed from the Devotini simulator with all the data required for the graft preparation available on the device (Figure 5b, c).

Video 1. Positioning prototype device.

Video 2. Positioning caliper.

Figure 3: Placement of device on a Devotini aortic root simulator. (a) Preparation of aortic root. (b) Alignment of subannular sutures. (c, d) Fixation and positioning of device.

Figure 4: Simulation of coaptation. (a, b) Appearance of aortic leaflets at varying diameters of circle. (c) Optimal coaptation.

Figure 5: The appearance of device at optimal coaptation and graft preparation. (a) Positioning of caliper. (b) Locked device at an optimal coaptation. (c) Marking and preparation of graft.

Ex vivo bovine heart study
A prototype device was designed according to the aortic root measurements of a 3-kg bovine heart (Figure 6a, b). The following steps were planned: (i) fixation of the device to the aortic root through the subannular sutures; (ii) b y a ttaching c ommissure holders to the aortic commissures, positioning of the device and aortic resuspension; (iii) u sing a n adjustable circle, the aortic coaptation simulation and through a stable (secured) adjustable circular frame, positioning of the aortic caliper perpendicular to the aortoventricular plane; (v) at an appropriate or inappropriate aortic coaptation created, water tests; and (vi) whether positioning of the device changes due to manipulations of the surgeon and gravity.

Figure 6: Implantation of the prototype device. (a) Three-dimensional modeling of device. (b) Prototype device appropriate for cattle"s heart. (c) Preparation of aortic root. (d) Fixation of device.

i) Preparation of the aortic root: A sinvalvesparing aortic root replacement procedures, the three sinuses of Valsalva were resected, and the coronary buttons were prepared. Resuspension sutures were attached to the aortic commissures. Subannular sutures were placed (Figure 6c).

ii) Fixation of the device to the aortic root: Each pair of the subannular sutures were passed through the corresponding suture holes on each leg. The device was then advanced to the aortic root under the guidance of the sutures (Figure 6d). These five pairs of suture lines were then drawn into the slit surgical tubes to fix the device.

iii) Positioning of the device and the resuspension of the commissures: Aortic commissures were attached to the commissure holders. The adjustable circle was allowed to move upward until a point where the commissures became suspended. The aortic valve now stood with its three-dimensional structure, just as it does during its native functioning. Positioning of the device did not change and remained stable with manipulations of the surgeon, or when the plane was changed (Video 3).

Video 3. Positioning prototype device and resuspension of the commissures.

iv) Simulation of aortic coaptation and positioning of the caliper: Leaflet coaptation was obtained by gradually narrowing the adjustable circular frame (Figure 7a, b, Video 4). The pins were, then, placed on the adjustable circle and their contact with the insertion and free margin of the cusps was observed. (Figure 7c). The plane provided by the adjustable circle maintained stable perpendicular positioning of the pins (Video 5).

Video 4. Simulation of coaptation.

Video 5. Positioning caliper.

Figure 7: (a, b) Simulation of aortic coaptation and (c) positioning of caliper.

v) Water test: When an appropriate aortic coaptation was obtained, water test was performed to test the resistance of coapting leaflets to water load (Video 6).

Video 6. Water test.

Discussion

Compared with either mechanical or biological valve replacements, valve-sparing aortic root replacement has proved to be advantageous owing to its noninferior, even superior, perioperative mortality and follow-up reoperation rates, as well as lower rates of bleeding, and prosthesis-related complications, and improved quality of life.[22,23]

The aortic root is a complex anatomic structure with its components: the aortoventricular junction or aortic annulus, aortic cusps, aortic commissures, aortic sinuses, and the sinotubular junction, each of which having crucial roles for competent aortic valve function, particularly coaptation.[5] All aortic root dimensions, including the effective height, have been shown to be highly patient-specific and correlated with the body weight and size, and body surface area.[20]

To date, the proponents of the reimplantation technique have made valuable contributions to the aortic root and valve geometry not only to enhance our insight into valve anatomy, but also to improve surgical outcomes of the patients with aortic root aneurysms. A particular attention has been drawn to various aspects of the dynamic anatomical geometry of the valve apparatus, including the effective height, geometric height, commissure height, coaptation height, annulus diameter, and cusp repair.[5,8,10,14,17,20]

The device proposed herein to be used for the reimplantation technique has one explicit advantage. All valve geometry to be reconstructed and repaired can be simulated on the device with all its elements - in particular, the commissures and the cusps with mutual geometrical relations in concrete values - before proceeding with final reimplantation procedures. Thus, all that is necessary can be clearly visualized in a manner whatever the configuration the surgeon prefers, particularly the creation of the effective height.

As many authors emphasize about the reimplantation technique, there is a demanding need for standardization, which has an adverse effect on the operative time, particularly for older patients.[22] With the help of this device, the applicability of the David procedure may easily be extended to elderly patients with decreased times of myocardial ischemia and cardiopulmonary bypass.

Another feasibility of the device is that any necessary intervention for appropriate leaflet coaptation (patch repair, prolapse repair, leaflet repair, subcommissural annuloplasty) can be made during the simulation of coaptation, even before the graft size has been chosen.

The proposed device was specifically designed for David I procedures, for aortic root aneurysms with or without aortic insufficiency.[24] Yet, the use of this device can be extended to include the David IV and V procedures in which a larger graft is used.[5,25] Similarly, by adding minor modifications to the device, it can be also helpful in repair of bicuspid valves.

Limitations
The device was primarily designed for patients having a normal or near-normal aortic annulus. David et al.[26] found a mean graft diameter of 30.8 mm (range, 26 to 34 mm) in 167 consecutive patients undergoing aortic valve-sparing operations for aortic root aneurysms. The mechanism of the device was designed based on the reported range of graft size. Therefore, in case of a larger annulus, the adjustment of the device along with simulation would not be possible before downsizing the annulus. The author recommends placing slit surgical tubes through some of the sutures placed without pledgets, particularly those close to the commissures.

In conclusion, valve-sparing aortic root reimplantation technique has taken a long way since its first introduction in 1992 by Tirone David. However, its success largely depends on the experience and skills of the surgeon, which still is a limitation to its wider use and adoption. As emphasized by most authors, technical standardization is indispensable and remains the weakest aspect of this operation. The device presented herein may be a strong candidate to overcome limitations at the very beginning of the David procedure, which in turn make the procedure more feasible and reproducible.

Declaration of conflicting interests
The author declared no conflicts of interest with respect to the authorship and/or publication of this article. All illustrations and device material were designed and created by the author.

Funding
The author received no financial support for the research and/or authorship of this article.

References

1) Chiang YP, Chikwe J, Moskowitz AJ, Itagaki S, Adams DH, Egorova NN. Survival and long-term outcomes following bioprosthetic vs mechanical aortic valve replacement in patients aged 50 to 69 years. JAMA 2014;312:1323-9.

2) Glaser N, Jackson V, Holzmann MJ, Franco-Cereceda A, Sartipy U. Aortic valve replacement with mechanical vs. biological prostheses in patients aged 50-69 years. Eur Heart J 2016;37:2658-67.

3) Goldstone AB, Chiu P, Baiocchi M, Lingala B, Patrick WL, Fischbein MP, et al. Mechanical or Biologic Prostheses for Aortic-Valve and Mitral-Valve Replacement. N Engl J Med 2017;377:1847-57.

4) David TE, Feindel CM. An aortic valve-sparing operation for patients with aortic incompetence and aneurysm of the ascending aorta. J Thorac Cardiovasc Surg 1992;103:617-21.

5) David TE. How I do aortic valve sparing operations to treat aortic root aneurysm. J Card Surg 2011;26:92-9.

6) Erasmi AW, Sievers HH, Bechtel JF, Hanke T, Stierle U, Misfeld M. Remodeling or reimplantation for valve-sparing aortic root surgery? Ann Thorac Surg 2007;83:S752-6.

7) Arabkhani B, Mookhoek A, Di Centa I, Lansac E, Bekkers JA, De Lind Van Wijngaarden R, et al. Reported outcome after valve-sparing aortic root replacement for aortic root aneurysm: A systematic review and meta-analysis. Ann Thorac Surg 2015;100:1126-31.

8) Marom G, Haj-Ali R, Rosenfeld M, Schäfers HJ, Raanani E. Aortic root numeric model: annulus diameter prediction of effective height and coaptation in post-aortic valve repair. J Thorac Cardiovasc Surg 2013;145:406-11.

9) Komiya T. Aortic valve repair update. Gen Thorac Cardiovasc Surg 2015;63:309-19.

10) Miller DC. Valve-sparing aortic root replacement: current state of the art and where are we headed? Ann Thorac Surg 2007;83:S736-9.

11) Beyersdorf F, Rylski B. Current state of the reimplantation technique (DAVID Operation): surgical details and results. HSR Proc Intensive Care Cardiovasc Anesth 2012;4:73-6.

12) Aicher D, Fries R, Rodionycheva S, Schmidt K, Langer F, Schäfers HJ. Aortic valve repair leads to a low incidence of valve-related complications. Eur J Cardiothorac Surg 2010;37:127-32.

13) Langer F, Graeter T, Nikoloudakis N, Aicher D, Wendler O, Schäfers HJ. Valve-preserving aortic replacement: does the additional repair of leaflet prolapse adversely affect the results? J Thorac Cardiovasc Surg 2001;122:270-7.

14) David TE. Sizing and tailoring the Dacron graft for reimplantation of the aortic valve. J Thorac Cardiovasc Surg 2005;130:243-4.

15) de Kerchove L, Boodhwani M, Glineur D, Noirhomme P, El Khoury G. A new simple and objective method for graft sizing in valve-sparing root replacement using the reimplantation technique. Ann Thorac Surg 2011;92:749-51.

16) Kunihara T. Toward standardization of valve-sparing root replacement and annuloplasty. Gen Thorac Cardiovasc Surg 2018;66:685-91.

17) Schäfers HJ, Aicher D. Root remodeling for aortic root dilatation. Ann Cardiothorac Surg 2013;2:113-6.

18) Koolbergen DR, Manshanden JS, Bouma BJ, Blom NA, Mulder BJ, de Mol BA, et al. Valve-sparing aortic root replacement?. Eur J Cardiothorac Surg 2015;47:348-54.

19) Settepani F, Cappai A, Raffa GM, Basciu A, Barbone A, Berwick D, et al. Cusp repair during aortic valvesparing operation: technical aspects and impact on results. J Cardiovasc Med (Hagerstown) 2015;16:310-7.

20) Bierbach BO, Aicher D, Issa OA, Bomberg H, Gräber S, Glombitza P, et al. Aortic root and cusp configuration determine aortic valve function. Eur J Cardiothorac Surg 2010;38:400-6.

21) Boodhwani M, de Kerchove L, El Khoury G. Aortic root replacement using the reimplantation technique: tips and tricks. Interact Cardiovasc Thorac Surg 2009;8:584-6.

22) Leontyev S, Schamberger L, Davierwala PM, Von Aspern K, Etz C, Lehmann S, et al. Early and late results after David versus Bentall procedure: a propensity matched analysis. Ann Thorac Surg 2019. pii: S0003-4975(19)31758-8.

23) Salmasi MY, Theodoulou I, Iyer P, Al-Zubaidy M, Naqvi D, Snober M, et al. Comparing outcomes between valve-sparing root replacement and the Bentall procedure in proximal aortic aneurysms: systematic review and meta-analysis. Interact Cardiovasc Thorac Surg 2019;29:911-22.

24) David TE, Feindel CM, David CM, Manlhiot C. A quarter of a century of experience with aortic valve-sparing operations. J Thorac Cardiovasc Surg 2014;148:872-9.

25) Miller DC. Rationale and results of the Stanford modification of the David V reimplantation technique for valvesparing aortic root replacement. J Thorac Cardiovasc Surg 2015;149:112-4.

26) David TE, Feindel CM, Webb GD, Colman JM, Armstrong S, Maganti M. Long-term results of aortic valve-sparing operations for aortic root aneurysm. J Thorac Cardiovasc Surg 2006;132:347-54.

Keywords : Aortic insufficiency, David procedure, reimplantation, standardization, valve-sparing aortic root replacement