Herein, we present the first successful SMFSG performed for celiac truncus in a Crawford type V aortic aneurysm patient in Türkiye.
SURGICAL TECHNIQUE
A 69-year-old Caucasian male patient with
abdominal pain was consulted by the Emergency
Department. The patient was in the American Society
of Anesthestiologists (ASA) Class III status and the
associated comorbidities were chronic obstructive pulmonary disease, hypertension, and hyperlipidemia.
After computed tomography angiography (CTA),
the patient was diagnosed with a Crawford type V
descending aortic aneurysm with a diameter of 7 cm
(Figure 1). Due to the patient"s comorbidities and ASA
class, open surgery was not an option.
Procedure planning and device sizing were performed using three-dimensional vascular imaging (RadiAnt DiCOM viewer, version 2021.2.64 bit). The diameter of the aortic visceral branches, the distance between the branches, and the clock position of each branch were identified over the aortic axis (Figure 2a, b). All these anatomic measurements are given in Table 1. Oversizing was planned for approximately 15% for both proximal and distal ends. Therefore, a tapered 34/28/160-mm thoracic stent graft (Ankura? TAA Stent Graft, Lifetech Scientific, Shenzhen, China) was selected for the procedure. Surgeon-modified fenestrated stent graft was planned for celiac truncus. The procedure was performed under general anesthesia in the hybrid operating room. The right femoral artery was surgically exposed and left percutaneous femoral artery access was obtained. After heparin administration, a pigtail catheter was inserted and parked between the thoracic 12th - lumbar 1st vertebra for direct visualization of the celiac truncus and the superior mesenteric artery (SMA) with the anteroposterior position. Following the preparation of the femoral access routes, the stent graft was unsheathed completely on the surgical back table in a sterile manner. The endograft covering material (expanded polytetrafluoroethylene [ePTFE]) was fenestrated upon the previous meticulous measurements performed by the CTA, using an 11-sized scalpel (Figure 3a). The unsheathing procedure, fenestration, and re-sheathing procedure took around 15 min. Approximately 1.0x1.2-cm single fenestration was made for celiac truncus by giving special attention not to damage the metal braces to preserve the endograft integrity and catching the perfect orientation. The diameter of the fenestration was larger than the celiac truncus which was 0.8x0.7 mm. Of note, while introducing the endograft, marker "8" should be on the greater curvature and seen as a line. Being able to read the marker "8" or "0" during fluoroscopy confirmed the positioning of the endograft (Figure 3b). Once the fenestration was ensured to be oriented toward the aortic target vessel, the deployment was performed. The mean arterial blood pressure was lowered to 70 to 80 mmHg during the deployment to optimize accuracy. The completion angiography showed the coverage of the aneurysm sac, patency of celiac truncus, and no endoleak and no need for balloon intervention (Figure 4b). At one-month of follow-up, computed tomography demonstrated a perfect positioning of the endograft, no presence of endoleak, and intact visceral branches (Figure 4c).
The SMFSG requires experience and a learning curve for accurate planning to position the fenestration. Aortic angulations and tortuosity, superior mesenteric, celiac, and renal artery distances are the most important criteria for fenestration orientation. Our first case was for Zone 2 left subclavian artery (LSA) endo-revascularization with back table fenestration nearly two years ago for subacute type B aortic dissection and was the first documented SMFSG for LSA in Türkiye.[7] For a successful procedure, the surgeon must be accustomed to the endograft used. Every brand may be used for this purpose; however, we preferred the Ankura™ TAA Stent Graft owing to its conformity. This endograft has an ePTFE, dual-membrane material for biocompatibility and durability with nitinol endoskeleton. Two different shaped radiopaque markers ("8" and "0") on each side of the proximal landing zone increase the accuracy and facilitate the orientation. The perfect positioning is ensured, when the markers look like a line. There is a connecting bar just distal to radiopaque marker "8" on the greater aortic curvature, and the bar avoids shortening and provides axial support as seen in Figure 3a. While restoring the graft to the sheet after fenestration is complete, special care must be taken not to rotate the graft. Also, the markers "8" and "0" may rotate depending on the aortic neck angle and tortuosity during deployment.
Fenestrated stent graft technology requires preoperative accurate measurements to ensure precise matching of the native vessel and the fenestration. Using this method, proper patient selection is the key to success. The anatomical limitations must be recognized, as the greatest attention should be given to the length between the celiac truncus and SMA.[8,9] In this case, our endograft was distally deployed at zero distance to SMA and the distance of SMA to celiac truncus (3.48 mm) was very suitable for this procedure. Despite the increasing experience, a margin of error in placing fenestrations always exists and the mismatch possibility may lead to complications. That is the reason why we fenestrate the endograft larger than the ostium of the celiac truncus. Such endovascular interventions should be performed, particularly in large-volume centers and in experienced hands. The structural and technical characteristics of the endografts used must be well known. A protective distance of at least 1 cm should be left before, after, or between fenestrations.
Poor or delayed blood flow of the branch vessel on completion angiography creates a requirement for stenting.[5] While entering this procedure, we planned not to stent the celiac truncus as it may be complicated due to the short body and the splitting of the truncus at a short distance. Without a complication, we were not willing to use any of it.
Achieving successful results with a modified device is a work of art; however, there are also some limitations. There is a risk of spoiling the endograft integrity by the fenestration and the absence of bench testing of the endograft before clinical use is not possible. Although a homemade device may not go through the same integrity or durability as an approved one, the series of SMFSG shows perfect durability and successful early and mid-term results.[5,6] L egal issues i nvolved w ith s urgeonmodified devices conclude that device modification is legal and unregulated in most instances.[10]
In conclusion, surgeon-modified fenestration stent grafting is a fast, widely applicable, and safe both in elective and acute settings, and is also readily available, inexpensive, and effective method of preserving celiac flow during thoracic endovascular aortic repair procedures without sophisticated instruments. Nonetheless, the durability concerns necessitate additional studies.
Patient Consent for Publication: A written informed consent was obtained from the patient.
Data Sharing Statement: The data that support the findings of this study are available from the corresponding author upon reasonable request.
Author Contributions: All authors contributed equally to the article.
Conflict of Interest: The authors declared no conflicts of interest with respect to the authorship and/or publication of this article.
Funding: The authors received no financial support for the research and/or authorship of this article.
1) Upchurch GR Jr, Escobar GA, Azizzadeh A, Beck AW,
Conrad MF, Matsumura JS, et al. Society for Vascular Surgery
clinical practice guidelines of thoracic endovascular aortic
repair for descending thoracic aortic aneurysms. J Vasc Surg
2021;73(1S):55S-83S. doi: 10.1016/j.jvs.2020.05.076
2) Leon LR Jr, Mills JL Sr, Jordan W, Morasch MM,
Kovacs M, Becker GJ, et al. The risks of celiac artery
coverage during endoluminal repair of thoracic and
thoracoabdominal aortic aneurysms. Vasc Endovascular
Surg 2009;43:51-60. doi: 10.1177/1538574408322655
3) Mehta M, Darling RC 3rd, Taggert JB, Roddy SP, Sternbach
Y, Ozsvath KJ, et al. Outcomes of planned celiac artery
coverage during TEVAR. J Vasc Surg 2010;52:1153-8. doi:10.1016/j.jvs.2010.06.105
4) Kuo HS, Tsai KC, Chen JS. Endovascular aortic arch
reconstruction with parallel grafts: A dilemma of excessive
endograft oversizing. Acta Cardiol Sin 2020;36:351-9.
5) Shu C, Fan B, Luo M, Li Q, Fang K, Li M, et al. Endovascular
treatment for aortic arch pathologies: Chimney, on-the-table
fenestration, and in-situ fenestration techniques. J Thorac Dis
2020;12:1437-48. doi: 10.21037/jtd.2020.03.10
6) Zhang L, Wu MT, Zhu GL, Feng JX, Song C, Li HY,
et al. Off-the-shelf devices for treatment of thoracic
aortic diseases: Midterm follow-up of TEVAR with
chimneys or physician-made fenestrations. J Endovasc
Ther 2020;27:132-42. doi: 10.1177/1526602819890107
7) İşcan HZ, Ünal EU. Surgeon-modified fenestrated stent graft
deployment in type B aortic dissection. Turk Gogus Kalp
Damar Cerrahisi Derg 2021;29:285-9. doi: 10.5606/tgkdc.
dergisi.2021.20641
8) Gökkurt Y, Yürekli İ, Çakır H, İner H, Yeşilkaya N, Karaağaç
E, et al. Mid- and long-term results in patients with thoracic
endovascular aortic repair. Cardiovasc Surg Int 2021;8:103-8.
doi: 10.5606/e-cvsi.2021.1141