Methods: Between June 2013 and April 2020, nine patients (8 males, 1 female; mean age: 47±11.9 years; range, 30 to 61 years) who underwent off-pump left ventricular assist device implantation and nine patients (8 males, 1 female; mean age: 47±11.4 years; range, 29 to 60 years) who underwent on-pump minimally invasive left thoracotomy were retrospectively analyzed. Postoperative outcomes and mid-term results of both groups were evaluated.
Results: Outflow graft was anastomosed to the ascending aorta with J-sternotomy in all patients. The median duration of intubation and intensive care unit stay were one (IQR: 1.5) day and eight (IQR: 6.5) days in the off-pump group, respectively and one (IQR: 0) day and seven (IQR: 7) days in the on-pump group, respectively. Intra-aortic balloon pump was needed during the weaning of cardiopulmonary bypass in one (11%) of the patients in both groups. Postoperative right ventricular failure was observed in two (22%) patients in the offpump group who were treated medically and recovered. There was no need for revision due to bleeding or postoperative extracorporeal membrane oxygenator implantation in either group. In the off-pump group, three patients underwent heart transplantation after median 854 (IQR: 960) days. Three patients died one month, two and four years after implantation. Three patients were still alive with left ventricular assist device and were being uneventfully followed for 365, 400, and 700 days after implantation.
Conclusion: Off-pump technique is safe and feasible option for implantation of left ventricular assist device via minimally invasive left thoracotomy.
In LVAD surgery and other cardiac interventions, there is an orientation toward minimally invasive techniques by its nature. The main factors that cause this orientation in heart failure surgery are the devices miniaturized and operations that can be performed via minimal incisions.[3-5] Moreover, minimally invasive LVAD surgery via left thoracotomy (MILT) is an increasingly important option owing to less trauma, blood loss, arrhythmogenic complications, and less intensive care hospitalizations.[6,7]
The tendency to off-pump surgery has become inevitable to reduce possible postoperative complications such as blood product use.[8] On-pump surgery has detrimental effects on coagulation system by activation of systemic inflammatory response.[9] Moreover, reducing the need for blood products also reduces the likelihood of patients encountering possible blood antigens, thereby reducing the risk of sensitization of transplant candidates.[10]
In the present study, we aimed to compare our off-pump LVAD implantation via MILT experience with on-pump surgery and to evaluate the impact on postoperative outcomes and follow-up.
Device
Continuous-flow, centrifugal pumps were used
for surgery. The HeartWare® (HVAD; Medtronic Inc., Framingham, MA, USA) and HeartMate 3®
(HM3; Abbott Inc., Chicago, IL, USA) systems and
their implantation have been previously described for
minimally invasive surgery and ascending/descending
aorta anastomosis.[4,5,11,12]
Surgical technique of minimally invasive
implantation of LVAD
After general anesthesia and single-lumen
endotracheal intubation, transesophageal
echocardiography probe was placed, and implantable
cardioverter defibrillator was switched off before
surgery. First, the LV apex was reached through
the left fourth or fifth intercostal space. Then, the
appropriate location for the inflow cannula of the LV
apex with transesophageal echocardiography (TEE)
was marked and four felt sutures were placed, one
for each quadrant and the sewing ring was sutured
to the LV apex with 3/0 polypropylene or polyester
stitches individually. Insufflation of carbon dioxide
(CO2) in the surgical field with a flow set at 2 to 4 bar
was initiated. The ascending aorta was exposed via
mini-J-sternotomy for outflow anastomosis. After
administration of unfractionated heparin to achieve
an activated clotting time of at least 300 sec, the
aorta was partially clamped and outflow anastomosis
was performed with a continuous polypropylene
5-0 suture in end-to-side fashion. The clamp was,
then, removed, allowing the blood to de-air the graft
and the graft was clamped again. The graft was
directed intrapericardially to the apex of LV. The
driveline was tunneled from the thoracotomy to the
right upper quadrant of abdomen. Outflow graft and
device were combined as suggested and de-airing
of the pump was repeated. With rapid ventricular
pacing, blood pressure was lowered briefly, and
coring was performed through the sewing ring.
Inflow cannula was passed through the sewing ring
by performing de-airing maneuver and the device
was secured. The speed of LVAD was gradually
increased according to the septum position. We
attempted to close the pericardium over the LVAD,
whenever possible. In patients with previous surgery,
a polytetrafluoroethylene membrane to avoid the
adhesion to the surrounding lung tissue was needed.
Full reversal of heparin by protamine infusion was
administered.
Statistical analysis
Statistical analysis was performed using the
SPSS version 15.0 software (SPSS Inc., Chicago,
IL, USA). Continuous variables were expressed
in mean ± standard deviation (SD) or median (interquartile range [IQR]), while categorical
variables were expressed in number and frequency.
The Shapiro-Wilk test was used, if a continuous
variable followed a normal distribution. The Levene
test and independent samples t-test were used to test
for significance between two independent groups with
normal distribution. The Mann-Whitney U test was used for continuous variables and chi-square test or
Fisher exact test for categorical variables. A p value
of <0.05 was considered statistically significant.
Preoperative pulmonary artery catheterization, hemodynamics, laboratory, and echocardiographic parameters are summarized in Table 1. The mean LV ejection fraction was 15±5.2% and 19±7.7% in the off-pump and on-pump groups, respectively (p>0.05). The mean values of right ventricular (RV) - fractional area change (FAC) and tricuspid annular plane systolic excursion (TAPSE) were 25±7.4% and 13.5±2.8 mm, respectively and 28±5.2% and 14.6±3 mm, respectively in the off-pump and on-pump groups (p>0.05). All other preoperative pulmonary artery catheterization, hemodynamics, laboratory, and echocardiographic parameters were similar in both groups (p>0.05). The HeartMate 3® in one (11%) patient and HeartWare® in eight (89%) patients in the off-pump group and the HeartMate 3® in four (44%) patients and HeartWare® in five (56%) patients in the on-pump group were implanted (Table 2).
Postoperative outcomes and follow-up
The mean duration of cardiopulmonary bypass
(CPB) time in the on-pump group was 63±22 min.
Three (33%) patients in the off-pump group and
five (55%) patients in the on-pump group needed
intraoperative red blood cell (RBC) replacement.
The median duration of intubation and intensive
care unit stay (ICU) were one (IQR: 1.5) days and
eight (IQR: 6.5) days in the off-pump group and one
(IQR: 0) days and seven (IQR: 7) days in the on-pump
group, respectively. Intra-aortic balloon pump (IABP)
was needed during the weaning of CPB in one (11%)
of the patients in both groups. Postoperative RV
failure (RVF) was observed in two (22%) patients
in the off-pump group who were treated medically
and recovered. There was no need for revision due to
bleeding or postoperative extracorporeal membrane
oxygenator (ECMO) support in either group. In the
off-pump group, three patients underwent heart
transplantation after median 854 (IQR: 960) days.
Three patients died one month, two and four years
after implantation. Three patients were still alive with LVAD and were being uneventfully followed
for 365, 400, and 700 days after implantation. In
the on-pump group, two patients underwent heart
transplantation 510 and 850 days after implantation,
respectively. Four patients died in one, four, five
and six years after implantation, respectively. Three
patients were still alive with LVAD and were being
uneventfully followed for five, five, and six years
after implantation, respectively (Table 2).
Left ventricular assist device implantation is a complex surgical procedure in patients with endstage heart failure. When these patients are compared with standard cardiac surgery patients, they have a higher surgical risk. Previous studies have shown that minimally invasive LVAD implantation is associated with less trauma, blood loss and infection, and shorter ICU stay.[7,13,14] Moreover, minimally invasive surgery is thought to prevent RVF, since the geometry of the pericardium does not deteriorate, the RV construction remains more stable.[15] Cardiopulmonary bypass has, itself, certain disadvantages such as activation of systemic inflammatory response and deleterious effects on the coagulation system.[9] The idea of offpump surgery is plausible, as these effects may disrupt hemodynamics and cause undesirable outcomes during or after surgery in patients with end-stage heart failure.
Off-pump LVAD implantation has no long-term outcomes in large series. In studies conducted with a small number of patients so far, the survival rate of off-pump LVAD implantation was 92% at one month and 85% at one year.[16,17] These results are consistent with those of standard LVAD implantation.[18] However, whether off-pump surgery is truly safe is still uncertain and needs to be investigated in randomized trials. Since off-pump surgery is performed in patients at high risk for CPB, these high-risk patients need long-term comparative results with those undergoing on-pump surgery.
One of the most detrimental complications of mechanical circulatory support, early cerebrovascular events that have not been documented after off-pump LVAD surgery in the studies, yet. During follow-up, number of emerging new cerebrovascular events are similar with on-pump surgery.[8] Despite the advantages of off-pump LVAD implantation, embolization of an atheroma is still a possible threatening issue during manipulations of ascending or descending aorta during outflow anastomosis.
One disadvantage of off-pump LVAD surgery is the LV cavity that cannot be examined thoroughly and may cause apical thrombi or trabeculae obstruct the inflow of the device. Thus, it is crucial to perform a precise intraoperative TEE and it is extremely important to investigate the heart cavities, if there is an obstacle that can clog the device. Moreover, without a circulatory assist of CPB, maintaining a stable hemodynamic status is challenging while sewing the ring, coring the apex, and securing the device. It is particularly important that the perfusion team, and the CPB circuit must be ready in the operation room, as hemodynamics of the patient with heart failure can be impaired at any time.
Conventional LVAD implantation under CPB via median sternotomy increases the need for blood transfusion.[19] Sensitization of the patients with LVAD implantation and concomitant transfusion have been reported to have a negative effect on survival and graft failure after heart transplantation.[20] In the study by Gaffey et al.,[21] it h as b een shown t hat increased transfusion rates may cause an increase in infection rates, which is associated with undesired events in heart transplantation. In Gregoric et al.'s[13] study, requirement of blood transfusion during surgery and within the early postoperative period in off-pump LVAD implantation was lower than the on-pump surgery. They also reported shorter duration of postoperative mechanical ventilatory support for patients in off-pump group. In the present study, although ICU stay seemed to be higher (mean 6 days), it was found to be lower compared to studies with sternotomy (13.1 days).[22]
Cardiopulmonary bypass has negative effects on platelet dysfunction, fibrinolysis, and degradation of coagulation factors.[23-25] This response is a pathway similar to the occurrence of an acute phase reaction seen in sepsis.[26] The activation of systemic inflammatory response is the cause of these adverse events.[9] Additionally, activation of these systems by LVAD may be exacerbated or aggravated the effects of CPB.[27] The off-pump surgery approach in patients with heart failure, which may be more affected by the negative effects of CPB than the normal patient population, may change the results positively.
Off-pump implantation of LVAD is a surgery that is thought to be associated with many risks, such as piercing and wearing the heart in common parlance, and that is a challenge for surgeons. However, while a patient with advanced heart failure has already risk for surgery, adding the adverse events of CPB on it can create negative effects for the mortality and morbidity. Off-pump approach is plausible in selected patient groups, and off-pump surgery is not an approach that should come to mind first by the teams at the beginning of the learning period. Surgeons must have been experienced in conventional technique of LVAD implantation, then perform MILT and off-pump surgery. Therefore, technique-related undesirable events may not be observed in the learning curve of off-pump surgery. In this approach, not only surgeons should not pay attention; but also, anesthesiologists and perfusionists should also be familiar with this technique and be prepared for its complications.
Nonetheless, this study has some limitations. First, the sample size is small. Second, although the data were collected prospectively, our study is limited by its retrospective design. Third, this is a single-center experience; therefore, outcome interpretation is limited by institutional bias.
In conclusion, off-pump left ventricular assist device implantation via minimally invasive left thoracotomy is safe and a feasible option. Our study results show that off-pump implantation of left ventricular assist device have similar outcomes with the on-pump group. Off-pump technique may be applied to prevent additional complications related to cardiopulmonary bypass. Avoiding cardiopulmonary bypass may not only prevent adverse events, but may also provide potentially better results at the time of heart transplant.
Ethics Committee Approval: The study protocol was approved by the Ankara City Hospital Ethics Committee (date: 03.03.2021, no: E1-1572-2021). The study was conducted in accordance with the principles of the Declaration of Helsinki.
Patient Consent for Publication: A written informed consent was obtained from each patient.
Data Sharing Statement: The data that support the findings of this study are available from the corresponding author upon reasonable request.
Author Contributions: Idea/concept: M.K., U.K., S.K.; Design: M.K., U.K., S.K., D.S.; Data collection: M.A., A.Y. C.K.; Analysis: D.S.; Writing article: M.K., U.K., S.K., D.S., M.A., Z.C.; Critical review: U.K., S.K., Z.C.
Conflict of Interest: The authors declared no conflicts of interest with respect to the authorship and/or publication of this article.
Funding: The authors received no financial support for the research and/or authorship of this article.
1) Kormos RL, Cowger J, Pagani FD, Teuteberg JJ, Goldstein
DJ, Jacobs JP, et al. The Society of Thoracic Surgeons
Intermacs database annual report: Evolving indications,
outcomes, and scientific partnerships. Ann Thorac Surg
2019;107:341-53. doi: 10.1016/j.athoracsur.2018.11.011.
2) Kirklin JK, Naftel DC, Pagani FD, Kormos RL,
Stevenson LW, Blume ED, et al. Seventh INTERMACS
annual report: 15,000 patients and counting. J Heart
Lung Transplant 2015;34:1495-504. doi: 10.1016/j.
healun.2015.10.003.
3) Tarzia V, Buratto E, Gallo M, Bortolussi G, Bejko J, Bianco
R, et al. Implantation of the HeartWare HVAD: From full
sternotomy to less invasive techniques. Ann Cardiothorac
Surg 2014;3:535-7. doi: 10.3978/j.issn.2225-319X.2014.08.12.
4) Kocabeyoglu SS, Kervan U, Sert DE, Unal EU, Demirkan
B, Guray Y, et al. Is it possible to implant HeartMate 3
less invasively? New pump, new approach. Artif Organs
2018;42:1132-8. doi: 10.1111/aor.13289.
5) Kocabeyoglu SS, Kervan U, Sert DE, Pac M. HeartMate 3
implantation via left thoracotomy with descending aortic
anastomosis. J Card Surg 2019;34:635-7. doi: 10.1111/
jocs.14097.
6) Hanke JS, Rojas SV, Avsar M, Haverich A, Schmitto JD.
Minimally-invasive LVAD implantation: State of the art.
Curr Cardiol Rev 2015;11:246-51. doi: 10.2174/1573403x110
3150514151750.
7) Dorken Gallastegi A, Hoşcoşkun EB, Kahraman Ü, Yağmur
B, Nalbantgil S, Engin Ç, et al. Long-term outcomes in
ventricular assist device outflow cannula anastomosis to the
descending aorta. Ann Thorac Surg 2022;114:1377-85. doi:
10) 1016/j.athoracsur.2021.08.071.
8) Strueber M, Meyer AL, Feussner M, Ender J, Correia JC,
Mohr FW. A minimally invasive off-pump implantation
technique for continuous-flow left ventricular assist devices:
Early experience. J Heart Lung Transplant 2014;33:851-6.
doi: 10.1016/j.healun.2014.05.016.
9) Clive Landis R, Murkin JM, Stump DA, Baker RA,
Arrowsmith JE, De Somer F, et al. Consensus statement:
Minimal criteria for reporting the systemic inflammatory
response to cardiopulmonary bypass. Heart Surg Forum
2010;13:E116-23. doi: 10.1532/HSF98.20101022.
10) Massad MG, Cook DJ, Schmitt SK, Smedira NG, McCarthy
JF, Vargo RL, et al. Factors influencing HLA sensitization
in implantable LVAD recipients. Ann Thorac Surg
1997;64:1120-5. doi: 10.1016/s0003-4975(97)00807-2.
11) Cheung A, Lamarche Y, Kaan A, Munt B, Doyle A, Bashir
J, et al. Off-pump implantation of the HeartWare HVAD
left ventricular assist device through minimally invasive
incisions. Ann Thorac Surg 2011;91:1294-6. doi: 10.1016/j.
athoracsur.2010.08.031.
12) Umakanthan R, Haglund NA, Stulak JM, Joyce LD, Ahmad R,
Keebler ME, et al. Left thoracotomy HeartWare implantation
with outflow graft anastomosis to the descending aorta:
A simplified bridge for patients with multiple previous
sternotomies. ASAIO J 2013;59:664-7. doi: 10.1097/
MAT.0b013e3182a708d1.
13) Gregoric ID, Radovancevic R, Akay MH, Jezovnik MK,
Nathan S, Patel M, et al. Short-term experience with offpump
versus on-pump implantation of the HeartWare left
ventricular assist device. ASAIO J 2017;63:68-72. doi:10.1097/MAT.0000000000000448.
14) Haberl T, Riebandt J, Mahr S, Laufer G, Rajek A, Schima
H, et al. Viennese approach to minimize the invasiveness of
ventricular assist device implantation?. Eur J Cardiothorac
Surg 2014;46:991-6. doi: 10.1093/ejcts/ezu051.
15) Pasrija C, Sawan MA, Sorensen E, Voorhees H, Shah A,
Strauss E, et al. Less invasive left ventricular assist device
implantation may reduce right ventricular failure. Interact
Cardiovasc Thorac Surg 2019;29:592-8. doi: 10.1093/icvts/
ivz143.
16) Moayedifar R, Riebandt J, Haberl T, Dimitrov K, Wiedemann
D, Schlöglhofer T, et al. Off-pump HeartWare HVAD
left ventricular assist device implantation. J Heart Lung
Transplant 2016;35:S319-S320.
17) Awad H, Abd El Dayem M, Heard J, Dimitrova G, Yu L, Sun
BC. Initial experience with off-pump left ventricular assist
device implantation in single center: Retrospective analysis. J
Cardiothorac Surg 2010;5:123. doi: 10.1186/1749-8090-5-123.
18) Tsiouris A, Paone G, Nemeh HW, Borgi J, Williams CT,
Lanfear DE, et al. Short and long term outcomes of 200
patients supported by continuous-flow left ventricular assist
devices. World J Cardiol 2015;7:792-800. doi: 10.4330/wjc.
v7.i11.792.
19) Aaronson KD, Slaughter MS, Miller LW, McGee EC,
Cotts WG, Acker MA, et al. Use of an intrapericardial,
continuous-flow, centrifugal pump in patients awaiting heart
transplantation. Circulation 2012;125:3191-200. doi: 10.1161/
CIRCULATIONAHA.111.058412.
20) McKenna DH Jr, Eastlund T, Segall M, Noreen HJ, Park
S. HLA alloimmunization in patients requiring ventricular
assist device support. J Heart Lung Transplant 2002;21:1218-24. doi: 10.1016/s1053-2498(02)00448-5.
21) Gaffey AC, Phillips EC, Howard J, Hung G, Han J, Emery
R, et al. Prior sternotomy and ventricular assist device
implantation do not adversely impact survival or allograft
function after heart transplantation. Ann Thorac Surg
2015;100:542-9. doi: 10.1016/j.athoracsur.2015.02.093.
22) Strueber M, O'Driscoll G, Jansz P, Khaghani A, Levy
WC, Wieselthaler GM; HeartWare Investigators. Multicenter
evaluation of an intrapericardial left ventricular assist
system. J Am Coll Cardiol 2011;57:1375-82. doi: 10.1016/j.
jacc.2010.10.040.
23) Belhaj A. Actual knowledge of systemic inflammation reaction
during cardiopulmonary bypass. Recent Pat Cardiovasc Drug
Discov 2012;7:165-9. doi: 10.2174/157489012803832784.
24) Momeni M, Carlier C, Baele P, Watremez C, Van Dyck
M, Matta A, et al. Fibrinogen concentration significantly
decreases after on-pump versus off-pump coronary artery bypass surgery: A systematic point-of-care ROTEM analysis.
J Cardiothorac Vasc Anesth 2013;27:5-11. doi: 10.1053/j.
jvca.2012.07.008.
25) Spiess BD. Transfusion of blood products affects outcome
in cardiac surgery. Semin Cardiothorac Vasc Anesth
2004;8:267-81. doi: 10.1177/108925320400800402.