Case Presentation
A 48-year-old man was admitted to our hospital with congestive heart failure. Exertional dispnea since childhood was noted in his past history. A transthoracic echocardiogram revealed left venticular hypertrophy and systolic dysfunction with an ejection fraction of 0.35, severe LVOTO (subaortic membrane, annular and valvular aortic stenosis), moderate aortic regurgitation, aneurysm of the ascending aorta, and calcified mitral valve stenosis. The peak LVOT and mitral valve gradients were 104 and 16 mmHg, respectively. At catheterization, left ventricle could not be entered. At the time of the operation, aneurysmal segment of the ascending aorta, stenotic aortic valves and a fibrotic subaortic membrane were resected, but LVOT was still severely stenotic and a 19 mm valve sizer could not be inserted. Through a standard left atriotomy a heavily calcified mitral valve was excised and replaced with a No. 27 St. Jude HP (St. Jude Medical, Inc, St. Paul, MN) mechanical prosthesis inserted. This procedure led to the further narrowing of the LVOT. The decision was made for an aortoventriculoplasty and aortotomy incision was extended into the right aortic sinus. The free wall of the right ventricle and the interventricular septum was incised as in a standard AVP procedure (Figure
1A). The aneurysmal portion of the ascending aorta was excised. The left and noncoronary sinuses of Valsalva were free of aneurysm and left in place (Figure
1B). One end of a 30 mm Dacron tubular graft was tailored as a patch for the repair of the interventricular septum and the aortic root (Figure
1C). This part of the graft was sutured to the interventricular septum with a continuous 3-0 polypropylene suture. This suture was interrupted at the aortic annulus, and a 27 mm valve sizer was inserted without any difficulty through the new annulus. A No. 27 St. Jude HP (St. Jude Medical, Inc, St. Paul, MN) mechanical prosthesis was secured to the aortic annulus and the patch with two-third of the sutures passing through the native annulus, and one-third from the patch. The portion of the patch distal to the annulus was sutured to the aortic sinus and aorta with a continuous 4-0 polypropylene suture. The distal end of the graft was anastomosed to the distal ascending aorta with a continuous 3-0 polypropylene suture (Figure
1D). The right ventricular outflow tract was closed with a separate triangle-shaped patch of Dacron (Figure
1E). Antegrade and retrograde blood cardioplegia was used for myocardial protection. The ischemic and cardiopulmonary bypass time was 186 and 200 minutes, respectively. The patient was extubated within 6 hours and discharged from the intensive care unit at 17
th hour, had an uneventful recovery and was discharged from the hospital at the 6th postoperative day. Echocardiographic peak gradients on the mitral and aortic valve measured 3 weeks after the operation were 6 and 12 mmHg, respectively.
Figure 1A. The aortic and the right ventricular incisions (solid line) and the incision of the interventricular septum (dashed line) are shown.
Figure 1B. The aneurysmal portion of the ascending aorta is excised, and the sinuses of Valsalva that are free of aneurysm are left in place.
Figure 1C. One end of a 30 mm Dacron tubular graft is tailored for patch closure of the interventricular septum and the aortic root.
Figure 1D. This part of the graft is sutured to the interventricular septum with a continuous 3-0 polypropylene suture and this suture is interrupted at the aortic annulus, and a mechanical prosthesis is secured to the aortic annulus and the patch with two-third of the sutures passing through the native annulus, and one-third from the patch. The portion of the patch distal to the annulus is sutured to the aortic sinus and aorta with a continuous 4-0 polypropylene suture. The distal end of the graft is sutured to the distal ascending aorta with a continuous 3-0 polypropylene suture.
Figure 1E. The right ventricular outflow tract is closed with a separate triangle-shaped patch of Dacron.