ISSN : 1301-5680
e-ISSN : 2149-8156
Turkish Journal of Thoracic and Cardiovascular Surgery     
Karotis arter stent planlamasında 3D baskılı anatomik karotis modellerin değerlendirilmesi
Hakan Göçer1, Ahmet Barış Durukan2, Osman Tunç3, Erdinç Naseri4, Ertuğrul Ercan5
1Department of Cardiology, Medical Park Uşak Hospital, Uşak, Turkey
2Department of Cardiovascular Surgery, Medical Park Uşak Hospital, Uşak, Turkey
3BTECH Company, METU Technocity, Ankara, Turkey
4Department of Cardiovascular Surgery, Afyon Park Hospital, Afyon, Turkey
5Department of Cardiology, Medical Park Izmir Hospital, Izmir, Turkey
DOI : 10.5606/tgkdc.dergisi.2020.18939

Abstract

Background: We aimed to investigate the potential role of threedimensional printed anatomical models in pre-procedural planning, practice, and selection of carotid artery stent and embolic protection device size and location.

Methods: A total of 16 patients (10 males, 6 females; mean age 75.6±4.7 years; range, 68 to 81 years) who underwent carotid artery stenting with an embolic protection device between January 2017 and February 2019 were retrospectively analyzed. The sizing was based on intraprocedural angiography findings with the same brand stent using distal protection device. Pre-procedural computed tomography angiography images used for diagnosis were obtained and modeled with three-dimensional printing method. Pre-procedural and threedimensional data regarding the size of stents and protection devices and implantation sites were compared.

Results: Measurements obtained from three-dimensional models manually and segmentation images from software were found to be similar and both were smaller than actually used for stent and embolic protection device sizes. The rates of carotid artery stenosis were similar with manual and software methods, but were lower than the quantitative angiographic measurements. Device implantation sites detected by the manual and software methods were different than the actual setting.

Conclusion: The planning and practicing of procedure with threedimensional models may reduce the operator-dependent variables, shorten the operation time, decrease X-ray exposure, and increase the procedural success.

Despite technical advancements in carotid artery stenting (CAS), there are still ongoing challenging issues.[1] Variants of carotid arterial system in terms of geometrical and spatial course and aortic arch variability are widely reported.[2-5] In addition, a wide range of instruments used during the procedure may confuse the operator when and where to use which. Therefore, 87 to 99% success rates have been reported.[2,3] Although pre-procedural planning is of utmost importance to overcome difficulties, the most optimal imaging modality for decision making is still unclear.[5,6]

Computed tomography (CT) and magnetic resonance imaging (MRI) are used for decision making in stent and embolic protection device (EPD) sizing at experienced centers. However, these modalities are mostly used for diagnosis, and sizing is solely based on angiography which is performed during implantation.[6-9]

Inappropriate sizing of stent and EPD (length and diameter), inappropriate design of stent, and inadequate coverage of the lesion by the stent may result in peri-procedural ischemic events.[10-14] Therefore, prespecification of preferences becomes a useful overall risk marker.[10,15,16]

Three-dimensional (3D) printing was first described by Charles Hull in 1986. It is a fabrication technique used to transform digital images into physical objects in a layer-by-layer approach where images from CT or MRI (digital imaging and communication in medicine [DICOM] files) are converted to a standard tessellation language (STL) file by software and finally printed by stereolithography (a special form of laser printing).[17-24] This software also enables us to make more accurate surfaceto- surface virtual measurements without printing. Although anatomical studies for educational purposes with 3D printing have been done in carotid arteries, there is no studies available regarding pre-procedural planning, practicing, and device selection.[14]

In this study, we aimed to evaluate optimal carotid artery stent and EPD sizes and implantation sites using 3D printed anatomical models and its software.

Methods

In this retrospective study, a total of 16 patients (10 males, 6 females; mean age 75.6±4.7 years; range, 68 to 81 years) who underwent CAS with EPD at Medical Park Izmir Hospital, Izmir, Turkey between January 2017 and February 2019 were included. The study group consisted of the patients stented throughout the study period. Having a high-risk surgery is the main indication for CAS. The procedures were performed by a single experienced interventionist. The Medtronic SpiderFX EPDs (Covidien, Mannsfield, MA, USA), sized 3.0-7.0 mm, were used as distal protection filters with the Medtronic Protégé RX (Medtronic Inc., Minneapolis, MN, USA) self-expanding straight and tapered stents of 6¥20 mm to 10¥60 mm in sizes. Stent and EPD size selection were based on twodimensional (2D) angiographic images, and quantitative angiography performed during implantation. Preprocedural CT images (Toshiba Aquilion One Vision Edition 640 slice Dynamic volume tomography; Canon Medical Systems, California, CA, USA) were used for diagnostic purposes (Figure 1). All procedures were performed by catheters and wires, which were suitable for aortic, carotid, and stenotic segment anatomies and through the femoral artery. The selection was based on the interventionist's spatial perception of 2D angiographic images and clinical knowledge during the procedure.

Figure 1: (a) Computed tomography angiographic (volumetric) anterior view of carotid arteries. (b) Computed tomography angiographic (volumetric) lateral view of carotid arteries. (c) 3D segmentation digital model of carotid arteries wall and lumens via mimics software. (d) 3D segmentation digital flow model of internal carotid narrowing via mimics software.
3D: Three-dimensional.

The patients" preoperative volumetric CT images were analyzed retrospectively and modeled with 3D printing where the DICOM files (Figure 1) on axial, coronal, and sagittal plan were converted to STL files (segmentation images) by Mimics Innovation Suite 21 Software (Materialise, Leuven, Belgium, CE 0120 Certification) (Figure 2). The arterial lumen and wall were created with data processing via manual smoothing and hollow command in design module (3-matic). Then, the STL files were exported to 3D printer (Formlabs Form 2, Formlabs Inc., MA, USA, and Laser Specifications: EN 60825-1:2007 certified) and printed using resin. The 3D models were treated with the wash and cure process and detached from the support points. All 3D models were prepared at the Middle East Technical University Technocity, BTECH Company facility on May 2019. The optimal stent length and diameter and EPD size were evaluated manually with a digital caliber on the physical model and 3D printing software on the digital model. These values were statistically compared with the actual used stent and EPD sizes from the patient records based on 2D angiography images (Figures 2 and 3). The percentage of carotid artery narrowing was calculated by the North American Symptomatic Carotid Endarterectomy Trial method in both physical model and digital model. This was compared to 2D angiographic measurements from quantitative angiography. To gain a better understanding of the spatial geometry of lesions, practice with 3D printed models was performed by the interventionist under X-ray in a non-sterile environment using different catheters and wires (Figures 4 and 5).

Figure 2: Measurements of Digital 3D printing models with 3D printing software (Mimics/3-matic) modules. (a) Proximal internal carotid artery inner diameter. (b) Distal internal carotid artery inner diameter. (c) Proximal internal carotid artery outer diameter. (d) Carotid bifurcation
3D: Three-dimensional.

Figure 3: Manuel measurements of 3D printed physical models with a digital caliper. (a) Distal internal carotid artery inner diameter measured with a digital caliper. (b) Proximal internal carotid artery inner diameter measured with a digital caliper. (c) Carotid narrowing lumen diameter with a digital caliper. (d) Distal internal carotid artery inner lumen diameter measured with a digital caliper.
3D: Three-dimensional.

Figure 4: Manual practice with different catheters, wires on 3D physical models in both non-sterile environment and under X-ray tube. (a) Left common carotid artery cannulation with guiding catheter in non-sterile environment. (b) Cannulation of common carotid artery and crossing the internal carotid narrowing via guide wire under angiographic guidance in anteroposterior view. (c) Left internal carotid artery cannulation with guiding catheter in non-sterile environment. (d) Cannulation of common carotid artery and crossing the internal carotid narrowing via guide wire under angiographic guidance in lateral view.
3D: Three-dimensional.

Figure 5: Manual practice with different size of embolic protection device, catheters, wires and stents on 3D physical models in both non-sterile environment and under X-ray tube. (a) Practice with different size of embolic protection device in non-sterile environment. (b) Practice with different size of embolic protection devices and stents in non-sterile environment. (c) Image of carotid stent and 3D digital model under angiographic image. (d) Practice with Protégé stent. (e) Practice with Protégé stent in antero-posterior view. (f) Practice with Protégé stent in lateral view.
3D: Three-dimensional.

A written informed consent was obtained from each patient. The study protocol was approved by the Medical Park Hospital Ethics Committee. The study was conducted in accordance with the principles of the Declaration of Helsinki.

Statistical analysis
Statistical analysis was performed using the PASW for Windows version 17.0 software (SPSS Inc., Chicago, IL, USA). Continuous variables were expressed in mean ± standard deviation (SD) or median (min-max), while categorical variables were expressed in number and percentage. The Friedman two-way analysis of variance (ANOVA) was used to test the significance difference. A p value of p<0.05 was considered statistically significant.

Results

Eight patients had a preoperative ischemic event. Pre-dilatation was performed in eight patients and post-dilatation was performed in three patients. There were four post-procedural ischemic events of which three were transient and one was permanent.

Measurements of each carotid stent diameter and length and EDP device size were similar for manual measurements via a digital caliper and segmentation images via the measurement module of Mimics software (p>0.05, Table 1). However, these measurements were lower than the actual implanted carotid stent diameter and length and EDP device based on 2D angiography images during implantation (p<0.001, Table 1). A bias toward the use of oversized stent and EPD were detected. Estimated carotid narrowing percentages were similar in both manual and software methods, but lower than the quantitative angiography measurements (p<0.001, Table 1).

Table 1: The actual and estimated (optimal) measurements of sizes of carotid stents and EPDs

Discussion

Although CAS is a widely performed interventional procedure, sizing and selection of devices are still ongoing issues. Proper sizing of device bears a high degree of correlation to success of procedure and decrease mortality and morbidity.[2,3] Undersizing may result in partial plaque coverage and inadequate revascularization, whereas oversizing may cause extensive endothelial damage and vasospasm.[6,11] Another important issue is distal protection device sizing where oversizing may result in vasospasm and undersizing may lead to embolization of the debris from the site of intervention, resulting in cerebrovascular events.[11,12] Placing the EPD at an adequate distance from the lesion and in a straight part of the distal lumen is also crucial to prevent devastating ischemic events and guarantee the procedure success.[2,3,6] In daily practice, the interventionist estimates the carotid stent diameter and length and EPD size using quantitative angiography images performed during implantation which have shortcomings due to the 2D nature of the imaging technique.[8] Evaluation of the lesion size for stent and distal vascular bed size for EPD with volumetric CT and MRI images give some advantages over 2D angiography. However, inadequate 3D reconstruction and poor resolution of images compared to 3D print software and printed models may decrease the success rates. In our study, we also demonstrated these shortcomings where sizing for the stent length and size and EPD size were similar for 3D printed and software-based measurements, but were smaller than the actual implanted sizes. We also documented a tendency for oversizing which may adversely affect the success rates as mentioned above.

The United States Food and Drug Administration approved 3D print software for medical purposes which provides more precise measurements not only with digitally structured 3D objects, but also 3D printed real models. The measurements are closer to reality and enable preoperative practice possible in non-sterile environment.[17,23] Therefore, preoperative planning of the procedure with 3D printed models is possible. This, in turn, leads to a decrease in the procedural time, radiation exposure of both the operator and patient, thereby, reducing complications, mortality, and morbidity.[14,17,23] The aortic arch type, crossing lesion in the tortuous vessel, inadequate catheter back-up, catheter-wire selection, and placing protection device can be also handled by pre-procedural practice and planning.[17] In this context, we attempted to demonstrate the importance of 3D printing in pre-procedural evaluation and documented deviances with 3D printed models and digitally created models via the print software.

The current European Society of Cardiology (ESC) guidelines define indications for carotid interventions and are based on stenosis percentage and symptomatology.[25] In particular, borderline lesions pose a confusion. Therefore, accurate measurements are vital. This study also revealed a tendency to overestimate stenosis percentage with 2D angiography. We believe that a more precise decision for identification of the lesion severity is possible with 3D printed models.

Device selection based on 3D printing measurements (manual or software) allows the operator making more precise decisions.[14,17,23] In our study, we demonstrated the importance of pre-procedural planning with 3D printing providing more accurate understanding of 3D spatial anatomy of vessel and lesion morphology. Moreover, preoperative printed models enabled performing preoperative practice in a non-sterile environment by increasing the operator self-confidence and giving him/her an opportunity to find out which maneuvers would be the best for the procedure. Although our study was mainly designed for finding out the optimal device size and placement sites, it also gave us an opportunity to practice on 3D models. If the method could have been employed pre-procedurally and the study was conducted prospectively, more accurate sizing may have been achieved and the outcomes of stenting may have been more favorable than the actual results. However, further large-scale, prospective, randomized-controlled studies are needed to comment on this subject and on possible comparison of the stent results employing this model with surgery.

Nonetheless, this study is retrospective in nature which may have led to bias. Prospective studies using pre-procedural planning with the model and comparing the results with actual results would be more valuable. In addition, the cost of the model is another limitation. Each model takes about 6 h to print which costs around ?300. Although the model may not be cost-effective currently, with wide use of the model, the cost is expected to decrease. Also, the method may prevent further interventions by enabling more accurate sizing.

In conclusion, our study demonstrated the importance of newly developing technologies and collaboration of medicine and engineering. It may be concluded that three-dimensional printing is a useful way for choosing the optimal size of the stent and embolic protection device and their placement sites in carotid interventions. Carotid artery stents and embolic protection device systems develop in parallel to industry and engineering. In addition, three-dimensional printing technology has been evolving with the advancement in software, printers, and printing materials. To decrease mortality and morbidity and to increase the procedural success, novel modalities such as three-dimensional printing can be integrated to interventional medicine via collaboration of engineers and interventionists. Nonetheless, to test these hypotheses, there is still a need for further prospective, randomized-controlled studies in this area. Based on our study results, we conclude that the importance of pre-procedural planning is of utmost importance and the use of three-dimensional printed anatomic models provide the interventionist practice before the procedure and allows him/her to become ready for possible issues and complications. In this way, more accurate sizing and localization may be possible which may increase the operator expertise and decrease complication rates and morbidity and mortality, eventually.

Declaration of conflicting interests
The authors declared no conflicts of interest with respect to the authorship and/or publication of this article.

Funding
All the expenses of the study were covered by the BTECH Company where the Author Osman Tunç is an employed engineer.

References

1) Zhang L, Zhao Z, Ouyang Y, Bao J, Lu Q, Feng R, et al. Systematic Review and Meta-Analysis of Carotid Artery Stenting Versus Endarterectomy for Carotid Stenosis: A Chronological and Worldwide Study. Medicine (Baltimore) 2015;94:e1060.

2) Cremonesi A, Castriota F, Secco GG, Macdonald S, Roffi M. Carotid artery stenting: an update. Eur Heart J 2015;36:13-21.

3) Morr S, Lin N, Siddiqui AH. Carotid artery stenting: current and emerging options. Med Devices (Auckl) 2014;7:343-55.

4) Schillinger M, Minar E. Matching carotid anatomy with carotid stents. J Cardiovasc Surg (Torino) 2008;49:723-7.

5) Wyers MC, Powell RJ, Fillinger MF, Nolan BW, Cronenwett JL. The value of 3D-CT angiographic assessment prior to carotid stenting. J Vasc Surg 2009;49:614-22.

6) Malik RK, Vouyouka A, Salloum A, Marin ML, Faries PL. Tips and techniques in carotid artery stenting. J Vasc Surg 2009;50:216-20.

7) Benndorf G, Strother CM, Claus B, Naeini R, Morsi H, Klucznik R, et al. Angiographic CT in cerebrovascular stenting. AJNR Am J Neuroradiol 2005;26:1813-8.

8) Remonda L, Senn P, Barth A, Arnold M, Lövblad KO, Schroth G. Contrast-enhanced 3D MR angiography of the carotid artery: comparison with conventional digital subtraction angiography. AJNR Am J Neuroradiol 2002;23:213-9.

9) Frölich AM, Pilgram-Pastor SM, Psychogios MN, Mohr A, Knauth M. Comparing different MR angiography strategies of carotid stents in a vascular flow model: toward stentspecific recommendations in MR follow-up. Neuroradiology 2011;53:359-65.

10) Conti M, Van Loo D, Auricchio F, De Beule M, De Santis G, Verhegghe B, et al. Impact of carotid stent cell design on vessel scaffolding: a case study comparing experimental investigation and numerical simulations. J Endovasc Ther 2011;18:397-406.

11) Wodarg F, Turner EL, Dobson J, Ringleb PA, Mali WP, Fraedrich G, et al. Influence of stent design and use of protection devices on outcome of carotid artery stenting: a pooled analysis of individual patient data. J Neurointerv Surg 2018;10:1149-54.

12) Kamenskiy AV, Pipinos II, Dzenis YA, Bikhchandani J, Gupta PK, Phillips N, et al. Effects of carotid artery stenting on arterial geometry. J Am Coll Surg 2013;217:251-62.

13) Blasel S, Hattingen E, Berkefeld J, Kurre W, Morawe G, Zanella F, et al. Evaluation of angiographic and technical aspects of carotid stenting with diffusion-weighted magnetic resonance imaging. Cardiovasc Intervent Radiol 2009;32:666-71.

14) Govsa F, Yagdi T, Ozer MA, Eraslan C, Alagoz AK. Building 3D anatomical model of coiling of the internal carotid artery derived from CT angiographic data. Eur Arch Otorhinolaryngol 2017;274:1097-102.

15) Gosnell J, Pietila T, Samuel BP, Kurup HK, Haw MP, Vettukattil JJ. Integration of Computed Tomography and Three-Dimensional Echocardiography for Hybrid Three- Dimensional Printing in Congenital Heart Disease. J Digit Imaging 2016;29:665-9.

16) Firouzian A, Manniesing R, Flach ZH, Risselada R, van Kooten F, Sturkenboom MC, et al. Intracranial aneurysm segmentation in 3D CT angiography: method and quantitative validation with and without prior noise filtering. Eur J Radiol 2011;79:299-304.

17) Russ M, O'Hara R, Setlur Nagesh SV, Mokin M, Jimenez C, Siddiqui A, et al. Treatment Planning for Image-Guided Neuro-Vascular Interventions Using Patient-Specific 3D Printed Phantoms. Proc SPIE Int Soc Opt Eng 2015;9417. pii: 941726.

18) Byrne N, Velasco Forte M, Tandon A, Valverde I, Hussain T. A systematic review of image segmentation methodology, used in the additive manufacture of patient-specific 3D printed models of the cardiovascular system. JRSM Cardiovasc Dis 2016;5:2048004016645467.

19) Wang DD, Gheewala N, Shah R, Levin D, Myers E, Rollet M, et al. Three-Dimensional Printing for Planning of Structural Heart Interventions. Interv Cardiol Clin 2018;7:415-23.

20) Deser SB, Demirağ MK, Kolbakır F. Does severe contralateral carotid artery stenosis affect the outcomes of carotid endarterectomy? Turk Gogus Kalp Damar Cerrahisi Derg 2019;27:35-42.

21) Lee M, Moharem-Elgamal S, Beckingham R, Hamilton M, Manghat N, Milano EG, et al. Evaluating 3D-printed models of coronary anomalies: a survey among clinicians and researchers at a university hospital in the UK. BMJ Open 2019;9:e025227.

22) Anwar S, Singh GK, Miller J, Sharma M, Manning P, Billadello JJ, et al. 3D Printing is a Transformative Technology in Congenital Heart Disease. JACC Basic Transl Sci 2018;3:294-312.

23) Kappanayil M, Koneti NR, Kannan RR, Kottayil BP, Kumar K. Three-dimensional-printed cardiac prototypes aid surgical decision-making and preoperative planning in selected cases of complex congenital heart diseases: Early experience and proof of concept in a resource-limited environment. Ann Pediatr Cardiol 2017;10:117-25.

24) Binder TM, Moertl D, Mundigler G, Rehak G, Franke M, Delle-Karth G, et al. Stereolithographic biomodeling to create tangible hard copies of cardiac structures from echocardiographic data: in vitro and in vivo validation. J Am Coll Cardiol 2000;35:230-7.

25) Aboyans V, Ricco JB, Bartelink MEL, Bjorck M, Brodmann M, Cohnert T, et al. 2017 ESC Guidelines on the Diagnosis and Treatment of Peripheral Arterial Diseases, in collaboration with the European Society for Vascular Surgery (ESVS). Rev Esp Cardiol (Engl Ed) 2018;71:111.

Keywords : Karotis arter, emboli önleme cihazı, Mimics yazılımı, baskı, stentleme, üç boyutlu
Viewed : 4950
Downloaded : 1188